![Ultrawave Q Series Operator'S Instruction Manual Download Page 21](http://html1.mh-extra.com/html/ultrawave/q-series/q-series_operators-instruction-manual_3534278021.webp)
21
Recording the data
System Requirements:
-
Windows 2000 SP4; XP SP2;Vista SP2’ 7
-
Microsoft Excel
-
Mac OS 10.1.2+
-
USB High Speed Hub Port (500mA)
-
One available USB port
To record the saved data:
Always ensure the memory card is inserted into the
Q-Series
prior to
turning the power on.
The results of each cycle will be saved to the card until the card is removed.
Once you are ready to record the information, remove the Memory card
from the
Q-Series
ultrasonic bath. Insert the supplied memory to USB card
reader into a USB port on the PC and insert the memory card into the
relevant port on the card reader.
There is a file on the memory card which is already set up to provide a
template for recording the data – Ultrawave.xls. Open this file and the data
will display in spreadsheet format.
The file will show information of every cycle run while the memory card has
been in place in the
Q-Series
ultrasonic bath.
To ensure that the next cycle information is recorded on the memory card,
switch off the
Q-Series
and reinsert the card before switching the bath on
again.
The data will only be saved to the card, if it is inserted before the machine
is switched on. If the Memory card is inserted when the
Q-Series
is already
switched on, the data will not be recorded.