Screen 6: Fluid level sensor
The
Q-Series
ultrasonic bath is fitted with a fluid level sensor to ensure it is
not under-filled prior to or during the cycle. This ensures that the optimum
cleaning process can occur and potential damage to the tank is minimised.
The level sensor is a resistive device, so will not detect fluids with very low
conductivity such as pure deionised (DI) water.
The level sensor can be disabled for specialist cleaning applications where
deionised (DI) water is used. There is a considerable risk of causing
damage to the tank if a minimum amount of water is not always retained
in the bath. Use with extreme caution.
Use the ON and OFF keys to set the Low Power function. When the desired
option is selected, use the MENU key to move to the next screen.
The fluid level setting will return to ON each time the unit is switched on.
WARNING
Setting the fluid level sensor to OFF means that the
Q-Series
ultrasonic
bath will operate as if the bath were full of fluid.
It is therefore the operator’s responsibility to ensure that the tank is filled
with sufficient water when the fluid level sensor is disabled.