Fault finding guidelines
This section is an aid to trouble shooting and should be used in conjunction with the relevant circuit
diagrams found at the rear of this manual. In each case the re calibration must be carried out after any
repair.
The following tables list some of the symptoms and the relevant circuit areas.
NOTE
Do not, under any circumstances attempt to repair the instrument whilst it is connected to
a patient.
Symptom
Suspect Circuit
Check
Unit does not switch ON
Remote switch on front panel J1 pin 1 goes low when
switch is depressed
U1 ON/OFF Control
Output goes low when switch
is depressed
Battery
J1 pin 1 > 7V
Battery Low LED on
permanently
Battery condition
J1 pin 1 > 7V
Replace battery
U3 Comparator
See test / calibration section
Excessive battery drain
U1 / Q2 ON/OFF Control
Q2 must switch when switch
is depressed.
Change battery
Ultrasound OK but no audio
Transducer position
Re position transducer
Crackling operation of volume
control VR1
Replace volume control
Broken conductor in cable
Check audio signal on pin 3
of U2
Headphone socket SK2
Audio signal on J3 pin 1
Replace loudspeaker
Unit unstable
Ultrasound transducer
Check output drive into
crystal is 1.5V pk/pk.
Check received signal in
transducer J1 pin 4, signal
should be a low frequency
audio signal approx. 2Vpk/pk.
Replace transducer.
©Ultrasound Technologies Ltd
Page 15 of 32
PD1 / FT120 service manual Issue 3 January 2007
Summary of Contents for PD1 series
Page 46: ...Q2 DRAIN Q2 GATE U3 PIN 1 U3 PIN 3 U1 PIN 1 UNIT SWITCHED ON U1 PIN 1 UNIT SWITCHED OFF...
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......