4.5 Calibration

Librae Incapacitance Tester is factory calibrated by Ugo Basile technicians. It is possible to 

override the original calibration if required by a customer quality policy (or other reasons).

For this purpose, Librae guides the user through a wizard calibration procedure. To run the 

procedure, press the CALIBRATION icon on the Home Page Menu (Figure 4.1.1).

NOTE: Before running the calibration wizard procedure, ensure that NO weight is applied on the 

food pads, AND a 100g calibrated weight is present. If necessary, Ugo Basile can supply a 100g 

calibrated weight. Refer to paragraph 7.0 for ordering details.

Figure 4.5.1 shows the calibration steps procedure. Follow the instructions on the screen to 

calibrate your Librae Incapacitance Tester.

Figure 4.5.1

DO NOT REMOVE

 the weight during the calibration procedure.

Follow the instructions from the calibration wizard.

After the execution of all steps, Librae Incapacitance Tester will be calibrated with the new 

values.

NOTE: Calibration is not necessary if it is not required by customer quality policy. The 

instrument is shipped fully calibrated.

NOTE: During the calibration procedure, Librae MUST be positioned on a stable workbench. 

In addition, Librae must be horizontal, level, and stationery. Avoid calibration if this is not 

possible..

NOTE: During calibration procedure, the values are stored ONLY after the confirmation of the 

last step. Before that, it is possible to escape the procedure just by pressing CANCEL button and 

NO DATA WILL BE STORED.

4.6 Set Autostart

Autostart is a unique Librae Incapacitance Tester feature. By selecting this operative mode, 

Librae helps the user to perform the experiment in a consistent way, reducing variability and 

user bias or intervention during measurement.

Before using this feature, it is important to set the parameters in order to take advantage of 

autostart.

Figure 4.6.1 shows the Autostart settings menu.

1.  Autostart Weight Threshold: this parameter sets the minimum weight the Librae MUST 

consider for animal. The percentage value from 10% to 90% is referenced to the animal 

weight (Refer to paragraph 4.3) 

2.  Autostart Stabilization Threshold: this parameter sets the percentage weight variation 

referenced to the animal weight (Refer to paragraph 4.3). Percentage value from 1% to 20%. 

3.  Autostart Sustain Time: this parameter sets the acceptable time for satisfying the previous 

two parameters. Numeric value from 200 msec to 5 sec.

Figure 4.6.1

To help the user understand how this feature works, an example is shown below.

Let’s assume the test has a mouse that weighs 50 g, the experiment duration is 15 seconds, and 

autostart mode is selected.

1.  Configure the SETUP menu (refer. to paragraph 4.3) as follows:

1.1 

Animal Max Weight: 60g;

1.2 

Measurement Duration: 15s;

1.3 

Start Mode; Auto.

2.  Press the SET AUTOSTART icon, and set the parameters as follows:

2.1 

Autostart Weight Threshold: 60%. With this value, Librae considers the presence 

 

of the animal ONLY IF MORE THAN 36g (60% of 60 grams) are measured.

2.2 

Autostart Stabilization Threshold: 10%. With this value, Librae considers that the 

 

animal is quiet and ready for experiment ONLY IF the NOISE on foot pads is 

 

LESS THAN 6g (10% of 60 grams). See Figure 4.6.2 for details.

2.3 

Autostart Sustain Time: 3s. With this value Librae NEEDS AT LEAST 3S in which the 

 

previous conditions are satisfied.

3.  Go to Home Page Menu, and press START button (Figure 4.6.2);

Summary of Contents for 47880-001

Page 1: ...Instruction manual Librae Incapacitance tester July 2020 revision 2 2 SKU 47883 47882 47885 47880 001 Pain Inflammation...

Page 2: ...organization Instruction Manual July 2020 Revision 2 2 MOD 13 Rev 1 CE CONFORMITY STATEMENT Manufacturer UGO BASILE srl Address Via G di Vittorio 2 21036 Gemonio VA ITALY Phone n 39 0332 744574 Fax n...

Page 3: ...d of this manual and also at ugobasile com Product features What is in the box Librae package contents SKU 47883 The Librae Incapacitance tester for mice 1 transparent plastic restrainer for mice 1 Pe...

Page 4: ...e the weight averaged over a few seconds is normally used as the data point For example the device provides an index of discomfort in the osteoarthritic knee normally expressed as the Left Right weigh...

Page 5: ...th the Librae touch screen you may notice a different less sensitive response You may need to press slightly harder or use a stylus pen on the touch screen This is because we employ a resistive touch...

Page 6: ...e you are welcome to contact our Service Department see paragraph 5 4 Customer Support specifying the serial number of your instrument 3 3 Safety Instructions The following guidelines must be followed...

Page 7: ...e manufacturer Servicing carried out by unauthorized personnel See also paragraphs 2 1 1 and 2 1 2 3 8 Connection Connect the mains cord between the power socket of the Librae and the power outlet pro...

Page 8: ...might take up to 60 seconds During this time the UB logo and product name will be displayed Figure 4 0 1 Figure 4 0 1 4 1 Home Page When the start up is finished the Home Page Menu is displayed Figur...

Page 9: ...l has entered the restrainer tilt the restrainer back down to the flat position 8 When ready press START button or pedal switch to run the experiment 6 1 If Autostart Mode is enabled wait for the inst...

Page 10: ...text field to enable a virtual keyboard Figure 4 2 2 This is an alphanumeric field enter a maximum of 12 characters Figure 4 2 2 Characters Virtual Keyboard Figure 4 2 3 Numeric Virtual Keyboard PROT...

Page 11: ...the restrainer and if the experiment is running correctly Figure 4 4 1 Ehorizontal lements displayed on the START page from left to right provide the following functionality 1 i INFO button This butto...

Page 12: ...start Autostart is a unique Librae Incapacitance Tester feature By selecting this operative mode Librae helps the user to perform the experiment in a consistent way reducing variability and user bias...

Page 13: ...tabilization AUTO icon is YELLOW When ALL the autostart requirements are completed the experiment automatically starts Figure 4 6 5 Figure 4 6 5 Autostart is waiting for stabilization time to elapse A...

Page 14: ...or update system icon refer to paragraph 4 1 1 5 Press Update button and wait for a reboot Figure 4 7 1 1 Now the Librae Incapacitance Tester is up to date An Erase DB could be necessary Refer to para...

Page 15: ...ton shows Figure 4 7 5 1 the results of the experiments stored inside Librae Incapacitance Tester Every result box contains 1 Date and time of the experiment 2 Treatment ID protocol stage and trial 3...

Page 16: ...gure 4 8 1 1 and Table 4 8 1 1 for connector pinout Figure 4 8 1 1 DB 15 Pin Signal Name Signal Type Description 1 Start Stop TTL OUT Start TTL High Level Stop TTL Low Level 2 Manual Autostart TTL OUT...

Page 17: ...not hesitate to contact our service department or our local distributor either directly of via our support page Before sending any instrument to our factory for repair please contact our logistics dep...

Page 18: ...paratus Thermal Place Preference TPP Test for Mice Rats Analgesy Meter the 1st and original Randall Selitto paw pressure test Other Ugo Basile products related to this device visit ugobasile com web s...

Page 19: ...torio 2 21036 Gemonio VA ITALY Tel 39 0332 744574 sales ugobasile com The 1st original Mouse RotaRod for motory coordination studies Fear Conditioning System ANYmaze GSM Grip Strength Meter for mice a...

Reviews: