Page 16
47200 Instruction Manual (Rev. 1)
5
5
.
.
Adjust the warning
noise threshold
value; it may be adjusted from 100gf to
1500gf in 10gf steps with a default value of
300gf
.
6
6
.
.
Adjust the force value (as a percentage of the full scale range) which
triggers
the start
of the measurement¸ it may be adjusted from 0.4% to 5.0%, in 0.1%
steps with a default value of
002.0%
.
7
7
.
.
Adjust the
maximum time
allowed for a measurement; adjustable from 10s to
30s in 1s steps with a default value of
15s
.
8
8
.
.
Use
Subject Numbers
or test only (
only Test
)
9
9
.
.
Store every measurement regardless, or only
good measurements
, i.e. ac-
cepted by the user (
Only good
)
1
1
0
0
.
.
Store the force
waveform or
only the
peak
value (
Waveform
)
1
1
1
1
.
.
Show
abbreviated configuration
settings on the
Measure menu (
Do not
)
1
1
2
2
.
.
Show the
offset value
before each measurement (
Do not
)
1
1
3
3
.
.
Adjust the
display brightness
,
from 1 to 10, with a default value of
10
.
1
1
4
4
.
.
Adjust Clock (see also paragraph 5.3-Calendar Clock)
To set current date and time, press
Change
; the
underlining shows the parameter being changed.
Decrease or increase the value by
Less
or
More
;
move to the following parameter (year, month, day,
hour, minute, second) by pressing
Next
. After
setting seconds, pressing Next will leave the menu
level.
Return to the
Main Menu
by pressing
Exit
; you will be asked whether to save the
changes to configuration, or not, see paragraph 4.5.
4.5.2
Restore Default Values
Pressing
Restore defaults
on the
Configure Menu
screen shows:
Pressing
Cancel
returns control to the
Main Menu
without altering the configuration settings or the da-
ta memory.
Pressing
Confirm
clears the data memory and restores the factory default
configuration values as shown above.