Page 13
47200 Instruction Manual (Rev. 1)
At this point, the user must choose whether to
Reject
or
Accept
the results of the test.
If the result is rejected, the user is given the option
to repeat the previous test (
Test
) or finish making
measurements (by pressing
Exit
,
which brings back
to the Measure Peak Force Menu).
Similarly, if the result is
Accepted
, the data
are stored in GSM’s memory and the user is
given the option to finish making measurements (
Exit
) or do another test (
Test
).
If the
Maximum measurement time
is exceeded
during the test, the test is terminated automatically.
Pressing
Begin again
effectively returns the user to
the first Measure Peak Force
screen, while
Repeat
test
allows the user to repeat the current test.
From the Measure Peak Force
screen, press
Next option
to View Acquired Data.
4.4.2
View Acquired Data
Selecting
View Acquired Data
from the
Measure
menu and pressing
Select this
dis-
plays briefly the total number of stored measurements, and then a screen similar to:
Which shows the first stored test (No. 1).
Press
Next
to view the next stored test (No. 2, in
the example),
Previous
to view the test before the
one being displayed, or
Exit
to return to option 1 of
the Measure menu (i.e. measure peak force).
4.4.3
Transfer Data to PC
This option allows data stored in GSM’s memory to be transferred to a PC. It requires
that GSM is connected to a PC via the USB cable supplied and that the PC is running the
DCA software provided with the instrument. Detailed operation is covered later in this
manual under
PC Software
, see paragraph 6.
If this option is selected when the instrument is not connected to a PC which is running
appropriate software, GSM displays
PC not available
for a few seconds and then returns
to the first option on the Measure menu.
4.4.4
Change Max Force
The next option enables the user to select the maximum force, among the available
ranges:
0-100gf
0-500gf
0-1500gf