Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
26
Pressing time (0 to 60 minutes)
This parameter is only active in the TEMPO and CADENCE mode.
Even if this value can be modified in the FIXED mode, it is not taken into account
Press level (1,5 to 5 bar)
Indicates the pressure in the injection cylinder. Adjustable from 1.3 to 5 bar, must adapt to different types
of cylinders for single or multiple point injection:
Cylinder type:
Press level
-
Cylinder Ø35 (100g) - 1 point injection Ø12/Ø13
1,5 bar
-
Cylinder Ø50 (200g) - 1 point injection Ø12/Ø13
1,5 bar
-
Cylinder Ø80 (500g) - 2 points injection Ø12
2,8 bar
-
Cylinder Ø80 (500g) - 3 points injection Ø12
4,2 bar
5.6.1.
PRESSING CYCLE PHASES
-
At the launch of the cycle, the operation of the cylinder is tested (1 movement back and forth)
-
At the end of the test, the chamber is closed and the temperature goes up according to the drying rate
until the preheating temperature is reached
If the preheating time = 0 or after preheating, an acoustic and optical signal will warn the user that the
cylinder can be placed on the pressing base
Press key
allows to open the chamber. The cylinder can be filled
-
After inserting the cylinder and press the key
again, the chamber closes and the cycle continues
-
Vacuum is on and the temperature goes up until it reaches the holding temperature
-
Injection starts at the end of the holding time
-
After pressing and vacuum flooding, the chamber opens and the tray stops at 1/3 of its total course
-
An acoustic and optical signal will warn the user that the cylinder can be removed
-
If press the key
, the tray is lowered