![UFSK 500 ECO User And Service Manual Download Page 36](http://html1.mh-extra.com/html/ufsk/500-eco/500-eco_user-and-service-manual_3240233036.webp)
Manual_500 ECO_SN0001-following_0520
500 ECO
www.ufsk-osys.com
User and Service Manual
Page 36 of 50
8.1.2 Daily maintenance routine
Function checks:
• Please check the stability of the treatment chair.
•
Please check the different adjusting options.
8.1.3 Annual maintenance routine
Annual maintenance routines may be performed only by licensed and qualified electricians pursuant to
DIN EN 62353.
Our after-sales service will be happy to provide quotations for our customers in the German-speaking countries.
Customer from foreign countries and non-German speaking countries are requested to contact our authorized
representative in their country for quotations. We will be happy to provide their addresses upon telephonic request.
8.1.4 Checks to be performed after changes of location or prolonged down times
General visual inspection:
• Damages of any nature to the visible chassis or frame elements, or the upholstery, or the manual keypad and
foot keypad and their respective cables; check for presence of all elements of the treatment chair.
• Damages to external cables: breaks, ruptures, cuts, crimps or bends, discolorations of the cable jackets,
abrasions, bulges, brittle areas, visible internal insulation or metallic wires.
• Damages to casings or coverings (manual or foot keypad, motor housing, plugs): Breaks, ruptures, cuts,
deformations, abrasion or signs of wear and tear
• Damages to the frame elements: Deformations, bulges, cracks or fractures, abrasion or signs of wear and tear
• Damages to plug-in type contacts: Bent pins, bent bases, damages to the contacts’ bushings or O ring collars
• Damages to the accessories attached to the product: Deformations, wear and tear, breaks or ruptures
• Check of the treatment chair for the presence of all its components
Function checks:
• Please check the function of the manoeuvring aid and lockability of the chair and check its stability.
Please watch out for conspicuous noises.
•
Please check the effectiveness of the strain relief
• Please check the installation of the cables
•
Please check the proper fit of the plug-type connections
•
Please check the proper fit of all connecting pins and their safety split pins;
watch out for any changes in the shape of the pins.
Site inspection:
• Please ensure the chair’s secure footing. It is not admissible to apply loads to the equipment on sloped surfaces
with an inclination angle
≥
10°.
• Please make sure of the appropriate safety clearances between the equipment and supply conduits, window
sills, etc.