7.2. Cleaning und disinfection of the plastic cover
For cleaning and surface disinfection of the chair‘s plastic cover sensitive cleaners and disinfectants with low-
alcohol concentration and/or suitable surfactants (compatible concentration!) or mixtures thereof are suitable.
Attention
Cleaning agents and disinfectants with an overly high alcohol content may lead to stress cracks in the
chair’s plastic cover and roughen its surface, especially after frequent repeated use. The same applies to
incompatible chemicals such as Acetone or quaternary ammonia compounds.
8.
Maintenance
If you have technical inquiries, require our customer service, or wish to order spare parts, please do not hesitate to
contact us at the address below:
UFSK-International OSYS GmbH
Kirchhoffstraße 1
D - 93055 Regensburg
Phone: +49 941 78862-15
Fax:
+49 941 78862-35
Mail:
Web:
Unauthorized alterations to the equipment and/or unauthorized opening of its electronic components without
prior consultation with the supplier will result in the automatic lapse of the manufacturer’s warranty and all product
liability.
The treatment chair has service life of approx. ten (10) years subject to average operation frequency and proper
maintenance and care by the user.
8.1. Maintenance routine
Caution
The following inspections and maintenance routines are not allowed to be performed on the device during
its use on patients and during the course of medical treatments.
Attention
As a general rule, only original spare parts shall be used for any repairs requiring the installation of new parts.
Identified defects need to be remedied immediately as the equipment will otherwise have to be p h a s e d
out.
8.1.1 Checks to be performed prior to every use
General visual inspection:
• Damages of any nature to the visible chassis or frame elements, or the upholstery, or the manual keypad and
foot keypad and their respective cables; check for presence of all elements of the treatment chair.
• Damages to external cables: breaks, ruptures, cuts, crimps or bends, discolorations of the cable jackets,
abrasions, bulges, brittle areas, visible internal insulation or metallic wires.
Site inspection:
•
Please ensure that the mandatory safety clearances are observed between the equipment and supply
conduits, window sills, etc.