background image

PAGE 3 OF 17 

0721 IH-4597

OPERATION CONTINUED

INITIAL CALIBRATION

When the scale is first installed, it should be calibrated 

to ensure accurate weighing results.

1.  Press and hold the 

ENTER/Cal

 button until CAL is 

displayed.

2.  When the button is pressed again and released, the 

display shows -C- while zero reading is stored. Next, the 

display shows the calibration weight value in grams.

3.  Place the specified calibration weight (sold separately) 

on the pan. See table below for appropriate weights.

4.  The display shows -C- while the reading is stored. 

When the display returns to the previous application 

mode, scale is ready for use.

WEIGHING MODE

Use this mode to measure the weight of a sample in the 

selected unit of measure.

1.  Press MODE/UNIT/Menu until the desired unit icon is 

displayed.

2.  Put an empty container on the pan, then press  

0/T/On/Off to tare the container weight.

3.  Add material to the container. The display shows the 

weight of the material.

PERCENT MODE

Use this mode to measure the weight of a sample as a 

percentage of a reference weight.

1.  Press MODE/UNIT/Menu until Set.rEF and the % icon 

are displayed.

2.  Press No to use the reference weight in memory and 

proceed to step 6.

3.  Press Yes to establish a new reference weight. Scale 

display will now blink until reference is stored.

4.  Put an empty container on the pan, then press  

0/T/On/Off to tare the container weight.

5.  Add the desired amount of material to the container, 

then press ENTER/Cal to store the displayed weight. 

The display shows 100%.

6.  Replace the reference material with new material. 

The display shows the percentage of the sample 

compared to reference weight.

SPECIFIC GRAVITY MODE

Use this mode to measure the specific gravity of a sample.

1.  Press 

MODE/UNIT/Menu

 until Set.rEF and the S icon are 

displayed.

2.  Press 

No

 to use the reference weight in memory and 

proceed to step 6.

3.  Press 

Yes

 to establish a new reference weight. Scale 

display will now blink until reference is stored.

4.  Put an empty container on the pan, then press  

0/T/On/Off

 to tare the container weight.

5.  Fill the container with water, then press 

ENTER/Cal

 to 

store the displayed weight. The display shows the water 

weight with the S icon blinking.

6.  Replace the water with the same volume of sample 

material, then press

 ENTER/Cal

. The display shows the 

specific gravity of the sample.

FLUID OUNCE MODE

Use this mode to measure the fluid ounce volume of a 

sample.

1.  Press 

MODE/UNIT/Menu

 until the FL icon is displayed.

2.  Put an empty container on the pan, then press  

0/T/On/Off

 to tare the container weight.

3.  Fill the container with sample material. The display 

shows the volume of the sample.

LEGAL FOR TRADE

When model V31XW3 (H-4597) or V31XW6 (H-4598) is used in 

trade or a legally controlled application, it must be set up, 

verified and sealed in accordance with local weights and 

measures regulations. It is the responsibility of the purchaser 

to ensure that all pertinent legal requirements are met.

SETTINGS

Before verification and sealing, perform the following steps:

1.  Set the security switch on the underside of the scale 

to the OFF position. (See Figure 2)

2.  Set LFT to OFF under Setup menu.
3.  Calibrate the scale.

4.  Enter the Units menu.

5.  Set the desired units of measure to ON.

6.  Exit the Units menu.

CALIBRATION WEIGHTS

Uline Model No.

H-4597

H-4598

Ohaus Model No.

V31XW3

V31XW6

Calibration Weights

3 kg

6 kg

Figure 2

Security Switch

ON

OFF

Summary of Contents for OHAUS VALOR 3000

Page 1: ...e in mode Stores reference sample in SG mode SECONDARY FUNCTION Long Press OFF Turns the indicator off UNIT Enters menu CAL Initiates span calibration process MENU FUNCTION Short Press No menu functio...

Page 2: ...Unit setting 3 Press No until the desired unit of measure is displayed 4 Press Yes to accept the new unit of measure ENABLE UNITS OF MEASURE Units of measure must be enabled in the menu to select fro...

Page 3: ...of the sample compared to reference weight SPECIFIC GRAVITY MODE Use this mode to measure the specific gravity of a sample 1 Press MODE UNIT Menu until Set rEF and the S icon are displayed 2 Press No...

Page 4: ...t Perform calibration Move scale to suitable location Cannot calibrate Unstable environment Incorrect calibration weight Move scale to suitable location Use correct calibration weight Cannot access mo...

Page 5: ...Capacity g 3000 6000 Repeatability Accuracy g 0 5 1 Linearity g 1 2 Approval Class III III Ingress Protection IP65 IP65 Weighing Units g kg lb oz lb oz Application Modes Weigh Percent Weighing Specif...

Page 6: ...rencia 100 en el modo Almacena una referencia de muestra en el modo SG FUNCI N SECUNDARIA Presi n larga OFF APAGADO Apaga el indicador UNIT UNIDAD Permite acceder al men CAL CALIBRACI N Inicia el proc...

Page 7: ...la unidad de medida deseada se muestre 4 Presiones Yes S para aceptar la nueva unidad de medida ACTIVE LAS UNIDADES DE MEDIDA Las unidades de medida deben ser activadas en el men para seleccionar del...

Page 8: ...a y luego oprima 0 T On Off 0 Tara Encendido Apagado para tarar el peso del contenedor 5 Agregue la cantidad del material deseada al contenedor y luego presione ENTER Cal ACEPTAR Calibraci n para alma...

Page 9: ...ma 1 12 Salga del men de configuraci n VERIFICACI N La autoridad local de pesos y medidas o un agente de servicio autorizado debe llevar a cabo un procedimiento de verificaci n SELLAR La autoridad loc...

Page 10: ...e transportaci n OVER Excedente El peso colocado sobre la bandeja excede la capacidad Quite peso de la bandeja UNDER Inferior La bandeja se quit durante el proceso de pesaje Vuelva a colocar la bandej...

Page 11: ...ara Hasta la capacidad por sustracci n Tiempo de estabilizaci n 3 segundos Requisitos de potencia 9 VDC hasta un m ximo de 1 2 A Adaptador AC se incluye o cuatro bater as C LR14 no se incluyen Calibra...

Page 12: ...nregistre 100 des r f rences dans un mode en Enregistre l chantillon de r f rence dans un mode SG FONCTION AUXILIAIRE Longue pression OFF ARR T teint l indicateur UNIT UNIT Acc de au menu CAL TAL Proc...

Page 13: ...ppuyez sur Yes pour accepter la nouvelle unit de mesure ACTIVER LES UNIT S DE MESURE Les unit s de mesures doivent tre activ es dans le menu afin de pouvoir tre s lectionn es partir du bouton MODE UNI...

Page 14: ...str e 4 Placez un contenant vide sur le bac puis appuyez sur 0 T On Off pour tarer le poids du contenant 5 Ajoutez la quantit d sir e de mat riel dans le contenant puis appuyez sur ENTER Cal pour enre...

Page 15: ...uvant sous la balance en position OFF arr t Voir Figure 2 2 Dans le menu Configuration r glez LFT en position OFF arr t 3 talonnez la balance 4 Acc dez au menu des unit s 5 R glez les unit s de mesure...

Page 16: ...ouillez le verrou de transport OVER Trop lev Le poids dans le bac est trop lourd Retirez du poids du bac UNDER Trop faible Le bac a t retir pendant le pesage R installez le bac Err 9 Les donn es d tal...

Page 17: ...liquide Plage de tare la capacit par soustraction Dur e de stabilisation 3 secondes Alimentation 9 VDC 1 2 A max Adaptateur c a fourni ou quatre piles C LR14 non fournies talonnage Num rique avec le p...

Reviews: