PAGE 4 OF 17
0721 IH-4597
7. Enter the Setup menu.
8. Set the desired Filter level.
9. Set the desired Power on Unit setting.
10. Set LFT to ON.
11. Set the Security Switch to ON. (See Figure 2)
12. Exit the Setup menu.
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. Refer to Figure 3 for sealing
methods.
Figure 3
WIRE SEAL:
Remove the spacers under the
cross-drilled screws. Reinstall
the screws and feed the wire
through the cross-drilled holes.
PAPER SEAL:
Place the paper seals over the
screws and sealing spacers.
OPERATION CONTINUED
TROUBLESHOOTING
OPERATING ISSUE
CAUSES
RECOMMENDATIONS
Cannot turn on.
No power to scale.
Verify connections and voltage.
Poor accuracy.
Improper calibration.
Unstable environment.
Perform calibration.
Move scale to suitable location.
Cannot calibrate.
Unstable environment.
Incorrect calibration weight.
Move scale to suitable location.
Use correct calibration weight.
Cannot access mode.
Mode not enabled.
Enter menu and enable mode.
Cannot access unit.
Unit not enabled.
Enter menu and enable unit.
Err 8.1
Pan has load during power on.
Transport lock is in locked position.
Remove weight from pan and re-zero.
Unlock transport lock.
Err 8.2
Pan was removed prior to power on.
Transport lock is in locked position.
Install pan and re-zero.
Unlock transport lock.
OVER
Weight on pan exceeds capacity.
Remove weight from the pan.
UNDER
Pan was removed during weighing.
Reinstall pan.
Err 9
Factory calibration data corrupted.
Contact Ohaus or an authorized
service agent. Perform service
calibration.
REF Err
Reference weight is too small.
Use larger sample.
LOWrEF
Reference weight is too low for
accurate percent weighing.
Continue to weigh with less accurate
results.
If the problem persists, or if the troubleshooting section does not resolve or describe your problem, contact
Uline Customer Service at 1-800-295-5510.