background image

PAGE 8 OF 8 

1214 IH-2289

trOuBleshOOting

OPERATING ISSUE

CAUSES

RECOMMENDATIONS

Unit will not turn on.

Power cord not plugged in or properly connected.

Power outlet not supplying electricity. 

Battery discharged.

Other failure.

Check power cord connections. Make sure power 

cord is properly plugged into the power outlet.

Check power source.

Replace batteries.
Service required.

Cannot zero the scale, or will not 

zero when turned on.

Load on scale exceeds allowable limits. 

Load on scale is not stable.

Load cell damage.

Remove load on scale.

Wait for load to become stable. 

Service required.

Unable to calibrate.

Lock Calibration Menu set to On.

LFT menu set to On.

Incorrect value for calibration mass.

Set Lock Calibration Menu to Off. 

Set LFT Menu to Off.

Use correct calibration mass.

Cannot display weight in 

desired weighing unit.

Unit not set to On.

Enable units in the Units Menu.

Cannot change menu settings. 

Menu has been locked.

Set selected menu to Off in the Lock Menu. 

Lockout Switch on the circuit board may need to 

be set to the Off position.

Error 8.1

Weight reading exceeds Power On Zero limit.

Remove load from scale. Recalibrate scale.

Error 8.2

Weight reading below Power On Zero limit.

Add load to scale. Recalibrate scale.

Error 8.3

Weight reading exceeds Overload limit.

Reduce load on scale.

Error 8.4

Weight reading below Underload limit.

Add load to scale. Recalibrate scale.

Error 8.6

Weight exceeds six digits. Display overflow.

Reduce load on scale.

Error 9.5

Calibration data not present.

Calibrate scale.

Battery Symbol Flashing

Batteries are discharged.

Replace batteries.

CAL E

Calibration value outside allowable limits.

Use correct calibration weight.

NO.SW

Attempting to exit the menu with the LFT setting ON 

and the security switch OFF.

Refer to Settings in Legal for Trade section, page 7. 

Set the security switch to the ON position.

REF. WT Err

Reference Weight too small. The weight on the 

platform is too small to define a valid reference 

weight.

Use a greater weight for sample.

π

  CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA

1-800-295-5510

uline.com

 If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.

maintenanCe

CautiOn! Disconnect the unit from the power 
supply before cleaning.

CLEANING

•  The housing may be cleaned with a cloth 

dampened with a mild detergent if necessary.

•  Do not use solvents, chemicals, alcohol, ammonia 

or abrasives to clean the housing or control panel.

Summary of Contents for H-2289

Page 1: ...xpanded resolution TARE Performs a tare operation Secondary Function Long Press OFF Turns the indicator off UNITS Changes the weighing unit MODE Allows changing the application mode Press and hold all...

Page 2: ...or End menu items pressing the YES button advances to the next menu while pressing the NO button returns to the top of the current menu CALIBRATION MENU Three calibration processes are available Zero...

Page 3: ...d by the user during the calibration procedure 1 When LINEAr is displayed press the YES button to initiate linearity calibration 2 The display flashes the full calibration point and calibration unit P...

Page 4: ...y The scale must be stable to accept zero operation manual tare When weighing an item that must be held in a container taring stores the container weight in memory 1 Place the empty container on the s...

Page 5: ...tton to clear the stored APW recalling a stored apw 1 Press the NO button to recall the existing APW NOTE Press the FUNCTION MODE button to temporarily display the APW value 2 The display shows the sa...

Page 6: ...Refer to Menu Navigation to enter settings NOTE The minus sign is used together with the first digit to show a negative value Editing the OVER setting 1 The display shows SEt HI Press the YES button...

Page 7: ...l pertinent legal requirements are met Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a...

Page 8: ...Error 8 2 Weight reading below Power On Zero limit Add load to scale Recalibrate scale Error 8 3 Weight reading exceeds Overload limit Reduce load on scale Error 8 4 Weight reading below Underload lim...

Reviews: