background image

PAGE 5 OF 8 

1214 IH-2289

Printing Data

Printing the displayed data to a printer or sending the 

data to a computer requires that the communication 

parameters in the Print and Communication Menu are set.

1.  Press the PRINT UNITS button to send the displayed 

data to the communication port (the Auto-Print 

Mode function must be off).

aPPliCatiOn mODes

1.  Press and hold the FUNCTION MODE button until the 

desired application mode appears. Only modes 

enabled in the mode menu will be displayed.

weighing

Place the item to be weighed on 

the scale. The illustration indicates a 

sample of 1.5 kg (gross weight).

nOte: Press the 

FUNCTION 

MODE 

button to temporarily 

display the weight in 10x 
expanded resolution.

Parts COunting

Use this mode to count parts of uniform 

weight. The Indicator determines the 

quantity based on the average weight 

of a single part. All parts must be 

uniform in weight for accurate measurements.

average PieCe weight (aPw)

When the FUNCTION MODE button is 

released, CLr.PW Pcs is displayed.

CLEARING A STORED APW

1.  Press the YES button to clear the stored APW.

RECALLING A STORED APW

1.  Press the NO button to recall the existing APW.

nOte: Press the 

FUNCTION 

MODE

 button to temporarily 

display the aPw value.

2.  The display shows the sample size 

PUt 10Pcs.

ESTABLISHING A NEW APW

1.  Press the NO button to increment 

the sample size. Choices are 5, 10, 

20, 50 and 100.

2.  To establish the APW, place the specified quantity 

of samples on the scale and press the FUNCTION 
MODE
 button to capture the weight.

3.  APW is displayed shortly followed 

by the APW value with the current 

unit of measure.

BEGIN COUNTING

1.  Place the parts on the scale and 

read the count. If a container is 

used, be sure to tare the empty 

container first.

PerCent weighing

Use this mode to measure the weight 

of a sample as a percentage of a 

reference weight.

referenCe weight

When the FUNCTION MODE button is 

released, CLr.rEF% is displayed.

CLEARING A STORED REFERENCE WEIGHT

1.  Press the YES button to clear the stored reference 

weight.

RECALLING A STORED REFERENCE WEIGHT

1.  Press the NO button to recall the existing reference 

weight.

nOte: Press the 

FUNCTION 

MODE 

button to temporarily 

display the reference weight.

OPeratiOn COntinueD

WEIGH

COUNt

.   1.500

 

kg

. 1.5000

 

kg

   0.123

 

kg

    0.123 

kg

PErCNt

CLr.rEF 

%

APW     

Pcs

AP        

Pcs

Summary of Contents for H-2289

Page 1: ...xpanded resolution TARE Performs a tare operation Secondary Function Long Press OFF Turns the indicator off UNITS Changes the weighing unit MODE Allows changing the application mode Press and hold all...

Page 2: ...or End menu items pressing the YES button advances to the next menu while pressing the NO button returns to the top of the current menu CALIBRATION MENU Three calibration processes are available Zero...

Page 3: ...d by the user during the calibration procedure 1 When LINEAr is displayed press the YES button to initiate linearity calibration 2 The display flashes the full calibration point and calibration unit P...

Page 4: ...y The scale must be stable to accept zero operation manual tare When weighing an item that must be held in a container taring stores the container weight in memory 1 Place the empty container on the s...

Page 5: ...tton to clear the stored APW recalling a stored apw 1 Press the NO button to recall the existing APW NOTE Press the FUNCTION MODE button to temporarily display the APW value 2 The display shows the sa...

Page 6: ...Refer to Menu Navigation to enter settings NOTE The minus sign is used together with the first digit to show a negative value Editing the OVER setting 1 The display shows SEt HI Press the YES button...

Page 7: ...l pertinent legal requirements are met Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a...

Page 8: ...Error 8 2 Weight reading below Power On Zero limit Add load to scale Recalibrate scale Error 8 3 Weight reading exceeds Overload limit Reduce load on scale Error 8 4 Weight reading below Underload lim...

Reviews: