19
Processor PCB assembly
Compressor
Power switch
Valve driver
PCB assembly
Figure 4: View of opened case
wires or tubes trapped, which will prevent the case from being closed properly. Turn
the controller over, and secure it with the four 3mm hexagon screws.
Perform the complete controller functionality tests (see Section 2.2 or 2.3).
Fit a new tamper label over the case retaining screw (see Figure 2).
4.2 REPLACING AN AIR HOSE SOCKET ‘O’ RING SEAL
This is done without opening the case.
Use a quarter to unscrew the
socket face
.
Remove the old
‘O’ ring seal.
Lubricate the new ‘O’ ring by lightly covering the tip of your forefinger with
‘O’ ring
lubricant
and rolling the ‘O’ ring between this finger and your thumb.
The most reliable way to insert a new ‘O’ ring is to put the ‘O’ ring onto the plastic
unit end connector. Put the connector and ‘O’ ring into the
socket body
, and turn the
socket face into the socket body. This will guarantee the ‘O’ ring is seated correctly.
When the ‘O’ ring is seated correctly, tighten the socket face with the quarter.
Perform the limited controller functionality test (see Section 2.2.9 or 2.3).
Power supply
assembly