Tuttnauer WTL198-0057 Operator'S Manual Download Page 8

Page 8 

Instructions for use

Record Table

Use the following table to record firmware updates. 

NOTE: 

manufacture firmware is printed on a label found 

at the bottom of the incubator and on its packaging.

Please, register your manufactory firmware in this manual 

for future reference: 

Manufacture Firmware:

Date

Version

Operator

Start-up

1-Place the auto-reader on a firm surface, free from 

vibrations, away from sunlight, currents of hot or cold air, 

chemicals and corrosive or flammable substances. Do not 

place the equipment in a way that disconnection of the 

plug from the power supply could be difficult. Leave a 

gap of at least 10 cm from the wall. Do not move the auto-

reader periodically or during its use. Connect the auto-

reader to a secure and stable electrical power source. 

Do not wet or heat the auto-reader. If liquid is 

spilled on the auto-reader, disconnect it and dry 

it immediately. Before turning on, verify that all 

reading positions are empty.  

To verify the proper functioning of the equipment, 

readout of unexposed BIs should be performed in every 

incubation position. The result from this readout should 

be positive. Follow the instructions of this guide in order 

to carry out readout.  

2- Turn on the incubator by connecting the end of the 

AC of the power supply to the power source and then 

connect the other end of the source (plug) on the rear of 

the unit. The equipment will start by running the last time 

and temperature set-up.

3-Synchronize the incubator with local time. See 

configuration methods in the 

Time setting mode

 section

All devices are manufactured with the following 

settings:

- Time zone: UTC +0:00

- Printing language: English 

If you want to change the time zone, refer to 

Modification 

of time zone

 section. If you want to change the print 

language, refer to 

Modification of print language

 section.

4- Check that required incubation temperature is correct. 

To change it, press button   

for 3 seconds. Lights 

corresponding to 37 ºC and 60 ºC will start blinking, 

indicating that the temperature program must be 

selected. Press the button   

to select the incubation 

temperature program, 37 ºC or 60 ºC. After 3 seconds, 

temperature modification will be accepted.

For 37 ºC, the auto-reader will automatically select 4 hs. 

final readout for all positions.

For 60 ºC the incubator allows choosing between 20 

minutes, 30 minutes, 1 hour, 2 hours or 3 hours final 

readout. It is possible to select an incubation program 

for each of the positions depending on the final reading 

characteristics of the indicators to be used.

Summary of Contents for WTL198-0057

Page 1: ...Rev 0 October 2019 Warning this product contains dry natural rubber...

Page 2: ......

Page 3: ...p 8 Remaining incubation time 10 Positive Control 10 Interpretation of results 10 Disposal 10 Audible alarm 10 Cancellation of a reading 11 Temperature monitoring 11 Thermal paper specifications 11 Re...

Page 4: ......

Page 5: ...Simultaneous incubation of BIs for EO processes control along with BIs monitoring the other processes is not possible Use description Composition Six incubation programs are available 4 hours at 37 C...

Page 6: ...emit electromagnetic fields Do not use the equipment on vibrating surfaces Do not pour any liquid inside Do not immerse into any liquid Use indoors only Disconnect the power cord before cleaning Do no...

Page 7: ...se of glass fragments generated after the glass ampoule inside the BI tube is crushed Cool the BI during the indicated time before crushing the ampoule Do not handle the BI excessively since this migh...

Page 8: ...s guide in order to carry out readout 2 Turn on the incubator by connecting the end of the AC of the power supply to the power source and then connect the other end of the source plug on the rear of t...

Page 9: ...rol every time a processed indicator is incubated For more information refer to the Positive Control section of this user guide 7 Place the rapid BI in a reading position close the protective cover an...

Page 10: ...sults are not valid until the biological indicator used as positive control shows a positive result A processed BI with a positive result indicates a failure in the sterilization process A negative re...

Page 11: ...3 mm The printer uses thermal paper rolls 1 To replace the paper roll pull the handle of the printer s door Open the printer s cover and remove the spent roll 2 Place the new paper roll with the outer...

Page 12: ...o reader and press the Clock Sync button the auto reader will be synchronized with the PC s clock Synchronization can only be done as long as there is not an ongoing reading firmware version available...

Page 13: ...at position which becomes disabled The incubation temperature is not stable Ongoing reading Theprintercoverisnottightlylocked Printer without paper Paper roll is placed in the wrong direction Check th...

Page 14: ......

Page 15: ......

Page 16: ......

Reviews: