Tuttnauer WTL198-0057 Operator'S Manual Download Page 12

Page 12 

This functionality allows changing the language in which 

the incubator printer renders the tickets.

To make this change, the device must be started in Setup 

mode (see 

Setup Mode

), then press the button       .

The current language and list of available languages will 

be printed to establish a reference. Press the button 

to scroll up and 

to scroll down. After 2 seconds, a 

ticket will be printed informing the selected language.

To confirm the change, press and hold the button 

for 

3 seconds. To discard the change, unplug the incubator 

from the power source.

Instructions for use

Setup Mode

Modification of time zone

Modification of printing language

The auto-reader has a setup mode from which, the device 

time zone and the print language can be set up. To enter 

Setup mode, turn on the incubator while holding down 

the  button   

. The red and green LEDs in the three 

positions will remain on, indicating that the device is in 

this mode. The printer will render a ticket indicating that 

it is in Setup Mode. 

To change the time zone and the print language, follow 

the instructions below. When finished, restart the 

incubator to exit Setup mode.

Time setting mode

Change the time using the firmware update utility 

Connect the auto-reader via the USB port to a PC and 

start-up the program. Select the Auto-reader and 

press the Clock Sync button, the auto-reader will be 

synchronized with the PC’s clock. Synchronization can 

only be done as long as there is not an ongoing reading.

firmware version available for the auto-reader; downloads 

it and installs it in a few seconds without the loss of any 

device data.

Update procedure

Before starting the process, turn off the incubator for five 

seconds, then turn it on and follow these steps:

1-Connect the incubator to a PC using a USB cable and 

make sure to have an internet connection.

2-Open the firmware update utility.

3-Select the device and press Start button to update.

4- Wait until the software indicates that the upgrade process 

is complete. The auto-reader prints an update confirmation 

ticket. If it was already up to date, it will print nothing.

NOTE: 

This process must be repeated every time a new 

firmware version is available, this will be informed in the 

website.

The time zone modification function allows changing the 

time zone of the device to suit the time zone of your country. 

To make this change, the incubator must be started in 

Setup mode (see 

Setup Mode

), then press the 

button. 

The current time will be printed to set a reference. 

To subtract hours, press the button   

as many times as 

necessary. To add hours, press the button   

as many 

times as necessary. After 2 seconds, a ticket will be printed 

informing the selected time. 

To confirm the change hold down the  button for 

3 seconds. To discard the changes, unplug the incubator 

from the power source.

Summary of Contents for WTL198-0057

Page 1: ...Rev 0 October 2019 Warning this product contains dry natural rubber...

Page 2: ......

Page 3: ...p 8 Remaining incubation time 10 Positive Control 10 Interpretation of results 10 Disposal 10 Audible alarm 10 Cancellation of a reading 11 Temperature monitoring 11 Thermal paper specifications 11 Re...

Page 4: ......

Page 5: ...Simultaneous incubation of BIs for EO processes control along with BIs monitoring the other processes is not possible Use description Composition Six incubation programs are available 4 hours at 37 C...

Page 6: ...emit electromagnetic fields Do not use the equipment on vibrating surfaces Do not pour any liquid inside Do not immerse into any liquid Use indoors only Disconnect the power cord before cleaning Do no...

Page 7: ...se of glass fragments generated after the glass ampoule inside the BI tube is crushed Cool the BI during the indicated time before crushing the ampoule Do not handle the BI excessively since this migh...

Page 8: ...s guide in order to carry out readout 2 Turn on the incubator by connecting the end of the AC of the power supply to the power source and then connect the other end of the source plug on the rear of t...

Page 9: ...rol every time a processed indicator is incubated For more information refer to the Positive Control section of this user guide 7 Place the rapid BI in a reading position close the protective cover an...

Page 10: ...sults are not valid until the biological indicator used as positive control shows a positive result A processed BI with a positive result indicates a failure in the sterilization process A negative re...

Page 11: ...3 mm The printer uses thermal paper rolls 1 To replace the paper roll pull the handle of the printer s door Open the printer s cover and remove the spent roll 2 Place the new paper roll with the outer...

Page 12: ...o reader and press the Clock Sync button the auto reader will be synchronized with the PC s clock Synchronization can only be done as long as there is not an ongoing reading firmware version available...

Page 13: ...at position which becomes disabled The incubation temperature is not stable Ongoing reading Theprintercoverisnottightlylocked Printer without paper Paper roll is placed in the wrong direction Check th...