Using the machine
E34: PRODUCT TIMEOUT The density of the chemical liquid could
Press RESET
be too high. Check.
If it persists, please contact
a technician.
E40: RELAY RES. H2O
There is a problem with the contactor
Press RESET
of the water heating elements.
If it persists, please contact
a technician.
E41: DRAIN TIMEOUT
The drain pipe could be obstructed or
Press RESET
crushed. Check it.
If it persists, please contact
a technician.
E42: AIR PRESSURE
The air pump could be
Press RESET
malfunctioning or there could be leaks
If it persists, please contact
of air in the ducts.
a technician.
E43: DRYING T.
The load may not be dry
Press RESET
because the preset minimum temperature
If it persists, please contact
was not reached.
a technician.
E44: PRE-WASH T.
A too high a temperature was detected
Press RESET
during the pre-wash phase.
If it persists, please contact
a technician.
E45: TANK T. LIMIT
A too high a temperature was detected
Press RESET
in the tank.
If it persists, please contact
a technician.
E46: AIR T. LIMIT
A too high air temperature
Press RESET
was detected.
If it persists, please contact
a technician.
E47: PHASE T. LIMIT
A too high a temperature was detected
Press RESET
during the phase in progress.
If it persists, please contact
a technician.
E50: PROBE 1 BROKEN The PT1000-1 working probe (tank)
Press RESET
could be faulty or disconnected.
If it persists, please contact
a technician.
E51: PROBE 2 BROKEN The PT1000-2 control probe (tank)
Press RESET
could be faulty or disconnected.
If it persists, please contact
a technician.
E52: AIR PROBE KO
The PT1000-3 probe (air) could be
Press RESET
broken or disconnected.
If it persists, please contact
a technician.
E53: TANK T. DIFF.
A temperature difference has been detected Press RESET
between the tank probes that is too high.
If it persists, please contact
a technician.
E62: BUS CABLE
A faulty connection has been detected
Press RESET
on the bus between the Micro-1 and Micro-2 If it persists, please contact
a technician.