44
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
5) Start the machine using the ON-OFF switch (1 Fig. 13).
6) At the beginning of the first cycle check the level probes of
the liquid suction tubes (detergent, neutraliser and/or oth-
er): pull one tube at a time out of the container and check if
the corresponding alarm flashes on the display, indicating
the need for liquid refilling.
7) After running 3-4 washing tests, clean the water filters
placed in the lower compartment of the machine (Fig. 9).
8) Check that the drain trap does not show water leaks and is
firmly secured to both the machine and the drain.
9) Check if the pumps correctly suck washing liquids. To do
this, check that the liquid rises along the tube connected to
them.
Fig. 13 - 1) ON-OFF switch. - 2) USB pendrive socket for PC
connection and to use the «Win Tuttnauer Service tool wash»
program. - 3) USB pendrive socket for inserting or extracting
the pre-filled programs. - 4) Rubber cap to close access to the
emergency release.
1
2
3
4