Digital C-FLUOR Submersible Probe

 

998-2131 

 

Rev. 1.0 

Page 13 

4.

 

Maintenance and Warranty

 

 

4.1 Maintenance 

4.1.1  Rinsing 

The C-FLUOR should be rinsed or soaked in fresh water following each 
deployment until it is completely clean again. 

4.1.2  Care for the bulkhead connector   

A light coat of Silicone spray should be used on the rubber of the male 
pins of the bulkhead to aid in sealing.  The manufacturer recommends 
3M™ Silicone Lubricant Spray or Loctite 8021 spray.   

Note:  You should avoid using silicone grease.  Do NOT use WD-40, 
it will destroy the connectors.

 

4.1.3  Care for the optics 

The optical window should be visually inspected after each deployment 
following a soaking in fresh water.  If cleaning is needed, use optical 
tissue to clean the window with soapy water. 
 

Note:  The C-FLUOR should 

NOT 

come in contact with any organic 

solvents (i.e. acetone, methanol) or strong acids and bases. 

 

4.2 Warranty Terms 

Turner Designs warrants the C-FLUOR and accessories to be free from defects 
in materials and workmanship under normal use and service for a period of 12 
months from the date of shipment from Turner Designs with the following 
restrictions: 

 

  Turner Designs is not responsible for replacing parts damaged by accident 

or neglect.  Your instrument must be installed according to instructions in the 
User’s Manual.  Damage from corrosion is not covered.  Damage caused by 
customer modification of the instrument is not covered. 

 

  This warranty covers only Turner Designs products and is not extended to 

equipment used with our products.  We are not responsible for incidental or 
consequential damages, except in those states where this limitation is not 
allowed.  This warranty gives you specific legal rights and you may have 
other rights which vary from state to state. 

 

  Damage incurred in shipping is not covered. 

 

 

Summary of Contents for C-FLUOR

Page 1: ...Digital User s Manual March 13 2020 P N 998 2131 Revision 1 0 TURNER DESIGNS 1995 N 1st Street San Jose CA 95112 Phone 408 749 0994 FAX 408 749 0998...

Page 2: ...ling Mode 9 3 1 2 Data Streaming Mode 10 3 13 Setting New Baud Rate 10 3 2 Calibration 11 3 2 1 Calibrating the Digital C FLUOR Probe 11 3 2 2 Restoring Factory Calibration 12 4 Maintenance and Warran...

Page 3: ...e return of an end of life product proceed as follows If you purchased your instrument through a Turner Designs Distributor please contact your local representative They will instruct you where to ret...

Page 4: ...re factory calibrated and a probe specific calibration certificate is included with each probe The certificate contains information that states the probe s minimum level of detection and range of conc...

Page 5: ...ification Letter 232 C FLUOR Without Connector Digital P N 2120 000 Identification Letter NC 232 Note No end cap is supplied for the C FLUOR Without Connector Digital see Section 2 5 For end cap speci...

Page 6: ...d SSS for in vivo Chlorophyll Phycocyanin Phycoerythrin Rhodamine Fluorescein P N 2100 900 Note The SSS can be used for both C FLUOR and Cyclops for the applications specified above Solid Secondary St...

Page 7: ...gramming Cable s USB plug into an available USB port on your computer and wait for the drivers to automatically install 2 Once USB drivers have installed download and install Digital C FLUOR s Softwar...

Page 8: ...then the instrument s detector is working 2 5 Digital C FLUOR with no connector Digital C FLUOR can be purchased without a 6 pin bulkhead connector or end cap If the No Connector version is purchased...

Page 9: ...are two modes that can be selected Polling and Data Streaming 3 1 1 Polling Mode Polling Mode is used when data storage or power is limited Power is supplied to the probe and a command is used to trig...

Page 10: ...change from red to green indicating the unit is streaming data Note Data rates faster than 1 reading per second will be updated at a rate of 1 reading per second in the Datapoint Value box To see the...

Page 11: ...rate Digital C FLUOR you ll need the following PC with Digital C FLUOR Software RS 232 Programming Cable P N 2120 160 12 Volt Power Supply P N 7000 941 Blank solution Standard solution of known concen...

Page 12: ...the Concentration Value you entered displayed in the Datapoint Value box confirming the calibration was saved 3 2 2 Restoring Factory Calibration To restore the factory calibration of your Digital C...

Page 13: ...n contact with any organic solvents i e acetone methanol or strong acids and bases 4 2 Warranty Terms Turner Designs warrants the C FLUOR and accessories to be free from defects in materials and workm...

Page 14: ...eturned free of charge for all customers in the contiguous continental United States For customers outside of the contiguous continental United States who purchased equipment from one of our authorize...

Page 15: ...Digital C FLUOR Submersible Probe 998 2131 Rev 1 0 Page 15...

Page 16: ...igital Data String RS 232 Temperature Range Ambient 0 to 50 C Water Temp 2 to 50 C Light Source Light Emitting Diode Excitation Wavelength Visible CHL RWT PC PE F UV CDOM fDOM Oil OB IR Turbidity Dete...

Page 17: ...Note Plastic may fluoresce and interfere with the sample s fluorescence 2 If using a glass container place the container on a non reflective black surface 3 Ensure that the probe is more than 3 inches...

Page 18: ...onnection Black 2 Supply Ground 0VDC PSU Ground Connection Orange 3 Receive Data RX Green 4 Transmit Data TX Blue 5 Digital Ground GND Yellow 6 N A Note Pin 6 is not used Pigtail Connector and Wire Co...

Page 19: ...ing sleeve Impulse P No MCDLS F C FLUOR Pins and Wire Color 3 4 2 1 5 6 3 4 2 1 5 3 4 2 1 5 3 3 4 4 2 2 1 5 5 Pigtail Connector and Wire Color 3 6 4 2 1 5 5 2 4 4 6 1 3 3 6 4 2 1 5 5 2 4 4 6 1 3 3 3 6...

Page 20: ...Stop bits 1 Flow Control None Once connected you can communicate using the list of commands below to setup collect information and program Digital C FLUOR Probes Command Name Note I Sensor Informatio...

Reviews: