Turbines Incorporated 

Cryogenic Flow Delivery System 

CDS1000 Manual (Rev E)

 

 

 

CDS1000 Page 

6

 

computing the fluid density for the measured temperature and adjusts the volumetric or mass 
delivery total. 

 

Simply stated, temperature compensation adds pulses to the pulse total when the detected 
temperature is colder than the products reference temperature and subtracts pulses when the 
product temperature is warmer than the reference temperature. The rate at which the pulses 
are added or subtracted is determined by the measured temperature departure from the 
products reference temperature. 
 

Modes of Operation: 

The 

CDS1000

 program offers 3 modes, Operating, Programming, and Prove. Any mode that 

can affect the calibration of the 

CDS1000

 is password protected. The 

CDS1000

 has a 2 tiered 

password level of security, Privileged and User/Weights and Measures. The Privileged 
password is maintained exclusively for Factory use and allows access to ALL menu options. 
The Weights and Measures password allows access to all displayed menu options except two, 
Initialization Options 5,8 Hardware test/calibrate and Loaner respectively. 

 

Operating Mode 

The Operating mode is the power up mode for the 

CDS1000

; all others must be accessed 

with a password. The Operating screen containing all delivery information is the normal 
display for this mode. The Detail screen is also accessible in this mode without a password. 

 

Programming Mode 

The Programming mode and associated menu’s permit the 

CDS1000

 to be configured to the 

operating conditions of the application. When accessed, the first Display is 

CDS1000 

Initialization Options menu containing 9 options. Options 2,3,5 and 8 are password protected. 
Options 5,8 require a Privileged password and are intended for Factory use only

.. 

 

Prove Mode 

The Prove mode permits convenient calibration of the trailer’s delivery system. This mode 
temporarily aborts certain safeguard features and therefore is password protected. To access 
this mode, depress the SETUP key while in Operating mode. The system will prompt for a 
password, either User/Weights and Measures or Privileged, to be entered. If an incorrect 
password is entered or the ESC key is depressed, the 

CDS1000

 will revert to the Operating 

mode. The Prove mode allows entering, changing or viewing the Kfactor. The Prove mode is 
retained until the 

CDS1000 

is either turned “Off” or no flow activity is detected for 5 

minutes. The Prove mode has the following operational characteristics: 
 

 The 

*

CDS1000

* message is replaced with the 

PROVER

 message on the operating screen 

  A password is not required to change the Kfactor 

  The pump cooldown timer is disabled (pump enabled) 

  The Totalizer can be reset regardless of the presence of flow 

 

Not a legal delivery” is displayed above the delivery totalizer and on a printed ticket 

Summary of Contents for CDS1000

Page 1: ...Box 933 15935 US Hwy 283 N Altus OK 73522 0933 Seneca 864 882 4544 http turbinesincorporated com Manual part number 900484 Revision E Copyright 2013 2016 Turbines Inc All rights reserved Information c...

Page 2: ...inted in the USA All Rights Reserved SAFETY INSTRUCTIONS The following instructions must be observed Every effort has been made to design and manufacture this instrument to be safe for its intended us...

Page 3: ...tallation 7 Operation 8 Programming 9 Specifications 13 Hardware 13 Software 13 Troubleshooting 14 Illustrations 15 Wiring 17 Return Policy 18 All Returns 18 Warranty Claim 18 Non Warranty Repair 18 E...

Page 4: ...ally exclude damage that has resulted from improper installation of improper maintenance of unauthorized change or modification to improper operation of and or misuse or abuse of the subject equipment...

Page 5: ...DS1000 The displayed total is corrected for temperature by sensing the resistance of the RTD temperature probe Delivery information consisting of 17 selectable parameters is transmitted via Bluetooth...

Page 6: ...s the normal display for this mode The Detail screen is also accessible in this mode without a password Programming Mode The Programming mode and associated menu s permit the CDS1000 to be configured...

Page 7: ...b Use at least 4 20 bolts with lock washers and or lock nuts 4 Install cables and connect to the rear panel Do not over tighten Connectors should be hand tight a Route cables and tie wrap excess so th...

Page 8: ...ss to PRINT only if printer is enabled To View uncompensated total Press to RESET totalizer to ZERO flow must be stopped Press to see unit DETAILS list and perform Battery Test Indicates Pump Cooldown...

Page 9: ...must be removed 1 Calibration 1 Average K factor 2 Linearizer Table 3 Linearizer Press 0 OFF 1 ON 2 Configuration 1 LOX 2 LIN 3 LAR 4 LCO2 5 LN2O 6 CCO2 7 MAPP 8 LPG Type value press ENT Not normally...

Page 10: ...Type value press ENT max 9999 Default 1234 for Turbines Bluetooth Printer Next Page Next Page 2 Totalize during alarm 3 Pump cooldown timer 4 Pump cooldown delay 5 Pump ON during alarm 6 Count while...

Page 11: ...rs 4 Print Example 005 19 Apr 11 14 29 Lin table 004 19 Apr 11 10 51 Factor 158 3444 003 16 Apr 11 15 54 Lin table 002 15 Apr 11 11 54 Factor 149 853 001 12 Mar 11 13 11 Lin table Press any key to con...

Page 12: ...port on the rear of the unit labeled OUTPUT The print timeout controls when the ticket totals are adjusted for a new ticket When the timer expires the value for start is set to the last value for stop...

Page 13: ...control self regulating non sweating temperature activated heater Power External 9 26 Vdc self resetting fuse 0 6A with heater OFF 1 6A with heater ON Transient protection Internal Rechargeable NiMH...

Page 14: ...from turbine meter Reconnect signal cable to coil b With a screw driver rapidly scratch coil below the threads on the bottom surface c Observe CDS1000 counts or signal light slightly illuminates or f...

Page 15: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 15 Illustrations...

Page 16: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 16...

Page 17: ...n Color Function Color 1 Pressure Input 4 20mA Brown Brown 1 Pump NC Brown 1 RTS N C 2 Pressure Input 4 20mA White White 2 Pump C White 2 CTS N C 3 Internal GND Blue Blue 3 Pump NO Blue 3 RX Red 4 Rat...

Page 18: ...anty Repair If a Product is returned for repair Turbines R D LLC will provide an estimate of the repair charges on a time and materials basis and the time required to make the repair If any repair est...

Page 19: ...rompted depress Enter and then ESC this places the CDS1000 in the prove mode operating screen The Prove mode allows repeated access to adjust the Kfactor while the cooldown timer flow inhibit of the r...

Page 20: ...e and remove the PEM With the CDS1000 powered up normally depress Setup and if not already in Prove mode enter your password Select Avg Kfactor by depressing 1 Example A new Kfactor is calculated to b...

Page 21: ...t page Power Off 11 12 2012 B Delete duplicate programming on Pg11 12 21 2012 C Incorporate Pressure PCO control feature and menu items 06 11 2014 D Added missing menu item 3 2 7 Pump off flow rate er...

Reviews: