Turbines, Inc. CDS1000 Installation And Operation Manual Download Page 8

Turbines Incorporated 

Cryogenic Flow Delivery System 

CDS1000 Manual (Rev E)

 

 

 

CDS1000 Page 

8

 

Operation 

 
 

 

 
 
 
 
 
 

 

Observe BLUE LED 
illuminated indicates 
unit is powered on 

Observe WHITE LED 
illuminated or flashing. 
Indicates unit is 
receiving signal from 
turbine meter. 

Observe "RED" 
LED "OFF". If 
flashing, Press "D" 
for alarm list. 

Press to enter 
"PROVE" mode. 
Password required.  
To exit "PROVE", 
power unit "OFF"

System power 
(ON/OFF) 

Alarm Quick list: 
scrolling icons 
indicating current 
alarms

 

Press to see 
list of "Current 
Alarms" 

Press to PRINT 
(only if printer is 
enabled) 

To View 
uncompensated 
total 

Press to 

RESET

 

totalizer to ZERO.  
(flow must be stopped) 

Press to see unit 

DETAILS

 list and 

perform Battery 
Test 

Indicates "Pump 
Cooldown Timer". Only 
appears during cooldown 
countdown and if enabled. 

Displays "Delivery 
total" (resettable) 

Grand totalizer 
(non-resettable) 

Temperature Indicator 

Rate Indicator 

Current 
Date/Time 

Current calibration 
(K) factor and last 
calibration date. 

Mounting Bracket 
P/N: 900040

NiMH Battery 
P/N: 900035 

POWER: Connect to 
vehicle battery 

SIGNAL: Connect to 
turbine flowmeter 

TEMP: Connect to 
temperature probe 

PRESS: Connect 
to 4-20mA output 

PCO: Connect to 
pump control 

OUTPUT: Connect 
to RS-232 printer 

Knob 
P/N: 900058

Summary of Contents for CDS1000

Page 1: ...Box 933 15935 US Hwy 283 N Altus OK 73522 0933 Seneca 864 882 4544 http turbinesincorporated com Manual part number 900484 Revision E Copyright 2013 2016 Turbines Inc All rights reserved Information c...

Page 2: ...inted in the USA All Rights Reserved SAFETY INSTRUCTIONS The following instructions must be observed Every effort has been made to design and manufacture this instrument to be safe for its intended us...

Page 3: ...tallation 7 Operation 8 Programming 9 Specifications 13 Hardware 13 Software 13 Troubleshooting 14 Illustrations 15 Wiring 17 Return Policy 18 All Returns 18 Warranty Claim 18 Non Warranty Repair 18 E...

Page 4: ...ally exclude damage that has resulted from improper installation of improper maintenance of unauthorized change or modification to improper operation of and or misuse or abuse of the subject equipment...

Page 5: ...DS1000 The displayed total is corrected for temperature by sensing the resistance of the RTD temperature probe Delivery information consisting of 17 selectable parameters is transmitted via Bluetooth...

Page 6: ...s the normal display for this mode The Detail screen is also accessible in this mode without a password Programming Mode The Programming mode and associated menu s permit the CDS1000 to be configured...

Page 7: ...b Use at least 4 20 bolts with lock washers and or lock nuts 4 Install cables and connect to the rear panel Do not over tighten Connectors should be hand tight a Route cables and tie wrap excess so th...

Page 8: ...ss to PRINT only if printer is enabled To View uncompensated total Press to RESET totalizer to ZERO flow must be stopped Press to see unit DETAILS list and perform Battery Test Indicates Pump Cooldown...

Page 9: ...must be removed 1 Calibration 1 Average K factor 2 Linearizer Table 3 Linearizer Press 0 OFF 1 ON 2 Configuration 1 LOX 2 LIN 3 LAR 4 LCO2 5 LN2O 6 CCO2 7 MAPP 8 LPG Type value press ENT Not normally...

Page 10: ...Type value press ENT max 9999 Default 1234 for Turbines Bluetooth Printer Next Page Next Page 2 Totalize during alarm 3 Pump cooldown timer 4 Pump cooldown delay 5 Pump ON during alarm 6 Count while...

Page 11: ...rs 4 Print Example 005 19 Apr 11 14 29 Lin table 004 19 Apr 11 10 51 Factor 158 3444 003 16 Apr 11 15 54 Lin table 002 15 Apr 11 11 54 Factor 149 853 001 12 Mar 11 13 11 Lin table Press any key to con...

Page 12: ...port on the rear of the unit labeled OUTPUT The print timeout controls when the ticket totals are adjusted for a new ticket When the timer expires the value for start is set to the last value for stop...

Page 13: ...control self regulating non sweating temperature activated heater Power External 9 26 Vdc self resetting fuse 0 6A with heater OFF 1 6A with heater ON Transient protection Internal Rechargeable NiMH...

Page 14: ...from turbine meter Reconnect signal cable to coil b With a screw driver rapidly scratch coil below the threads on the bottom surface c Observe CDS1000 counts or signal light slightly illuminates or f...

Page 15: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 15 Illustrations...

Page 16: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 16...

Page 17: ...n Color Function Color 1 Pressure Input 4 20mA Brown Brown 1 Pump NC Brown 1 RTS N C 2 Pressure Input 4 20mA White White 2 Pump C White 2 CTS N C 3 Internal GND Blue Blue 3 Pump NO Blue 3 RX Red 4 Rat...

Page 18: ...anty Repair If a Product is returned for repair Turbines R D LLC will provide an estimate of the repair charges on a time and materials basis and the time required to make the repair If any repair est...

Page 19: ...rompted depress Enter and then ESC this places the CDS1000 in the prove mode operating screen The Prove mode allows repeated access to adjust the Kfactor while the cooldown timer flow inhibit of the r...

Page 20: ...e and remove the PEM With the CDS1000 powered up normally depress Setup and if not already in Prove mode enter your password Select Avg Kfactor by depressing 1 Example A new Kfactor is calculated to b...

Page 21: ...t page Power Off 11 12 2012 B Delete duplicate programming on Pg11 12 21 2012 C Incorporate Pressure PCO control feature and menu items 06 11 2014 D Added missing menu item 3 2 7 Pump off flow rate er...

Reviews: