12
13
■
Operation maintenance:
■
Note
●
Air Turbine and Contra Angle
:
Remove the test bur from the handpiece.
●
Straight Handpiece
:
Insert the supplied test bur into the handpiece and locking.
(5) will show on LCD Display by pressing shortly. Press / /
to select handpieces which are connected to run optional auxiliary rotation.
(6) will show on LCD Display by pressing shortly.
■
Compressed air setting:
Adjust air pressure by turning between 4.0-6.0 bars. Recommend to set air pressure at
4.0bars for TTcare 100+. Save and leave functional setting mode by pressing shortly.
6. Maintenance and Replacement Parts
■
Replace Fuse
(1) Press fuse holder by slotted wrench to loose the holder.
(2) Remove fuse holder.
(3) Replace new fuse (250V2A Slow-Blow type) into fuse holder
(4) Insert fuse holder.
Fuse
Fuse holder
■
Oil amount adjustment :
(1) Indicator lights and are on when the Lubrication Unit is turned on. Press
for three seconds until it enters Oil amount adjustment mode, LCD will display:
(2) There are 11 adjustable oil amount from PU-00 to PU-10. Press for increasing or for
decreasing. PU-00 is the default setting, PU-10 has the largest oil amount.
Choose the PU value required, press button to complete setup. Lubrication unit will
automatically reboot after setting adjustments.
(1)
Chuck maintenance:
Turn on the power switch. Open the front door.
will show on LCD Display during process.
Place the chuck side of head to the chuck nozzle.
Press to begin cleaning and lubricating the chuck .
A beep sound meaning the procedure is completed.
(2)
Handpiece maintenance:
Turn on the power switch.
Open the front door.
Connect the handpiece to its coupling with the chuck facing back.
Make sure the handpieces are connected to the adapter until it clicks into place.
Close the front door.
will show on LCD Display during process.
Press / / to select handpieces which are connected.
Press to lubricate the handpieces in sequence order.
A beep sound meaning the procedure is completed.
(1) will show on LCD Display meaning compressed air is on.
Turn compressed air On/Off by pressing .
(2) will show on LCD Display by pressing shortly.
Maintenance time setting:
Select 15 sec by pressing /Select 20 sec by pressing / Select 30 sec by pressing .
(3) will show on LCD Display meaning buzzer sound is on by pressing shortly.
Select buzzer sound On/Off by pressing .
■
Note
Pressing any button or opening the front door will immediately stop the maintenance procedure.
■
Functional setting:
Select functional setting mode by pressing for approx 2 sec.
(4) will show on LCD Display meaning auxiliary motor is on by pressing
shortly. Select Auxiliary motor On/Off by pressing .