Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
The TTcare 100+ is intended for use in the electromagnetic environment specified below.
The customer or the user of the TTcare 100+ should assure that it is used in such an environment.
Guidance and manufacturer’s declaration-electromagnetic immunity
Electrostatic
discharge(ESD)
IEC 61000-4-2
Floors should be wood, concrete or
eramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Electrical fast
transient/burst
IEC 61000-4-4
Mains power quality should be that
of a typical commercial or hospital
environment.
± 2kV for power
supply lines
Not applicable
± 2kV for power
supply lines
± 1kV for
input/output lines
Surge IEC 61000-4-5
Mains power quality should be that
of a typical commercial or hospital
environment.
± 1kV differential
mode
Not applicable
± 1kV line(s) to
line(s)
± 2kV line(s) to
earth
The TTcare 100+ power frequency
magnetic fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
3 A/m
3 A/m
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
TTcare 100+ requires continued
operation during power mains
interruptions, it is recommended
that the TTcare 100+ be powered
from an uninterruptible power
supply or a battery.
<5% UT(>95%
dip in UT) for 0,5
cycle 40%
UT(60% dip in
UT) for 5 cycles
70% UT(30% dip
in UT) for 25
cycles <5%
UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip
in UT) for 0,5 cycle
40% UT(60% dip
in UT) for 5 cycles
70% UT(30% dip
in UT) for 25
cycles <5%
UT(>95% dip in
UT) for 5 s
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
■
Note:
UT is the a.c. mains voltage prior to application of the test level.
70
71
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance
The TTcare 100+ is intended for use in the electromagnetic environment specified below.
The customer or the user of the TTcare 100+ should assure that is used in such and environment.
Guidance and manufacturer’s declaration-electromagnetic immunity
Portable and mobile RF communications
equipment should be used no closer to any
part of the TTcare 100+ including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Conducted RF
IEC 61000-4-6
3 Vrms
3 Vrms
150 KHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
3 V/m
80MHz to 2,5 GHz
■
a:
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the TTcare 100+ is used
exceeds the applicable RF compliance level above, the TTcare 100+ should be observed to
verify normal operation. If abnormal performance is observed, additional measures my be
necessary, such as re-orienting or relocating the TTcare 100+.
■
b:
Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 3 V/m.
■
Note1:
At 80 MHz and 800 MHz, the higher frequency range applies.
■
Note2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.