16
17
7. Maintenance and Inspection
8. Trouble Shooting and Error Code
■
Trouble Shooting
Breakdown
Reason
Solution
Main power lamp does not light up
Does not start when Start Switch is
pressed.
(1) Double beep sound
(2) Spray can Empty Lamp is
blinking
Unusual odor.
Surfaces are unusually dirty.
Cannot install spray can.
Oil leaks from spray can.
Power supply cord not properly
plugged into supply socket
Power supply cord not properly
plugged into TT
Front door may be open
Empty can
Can is not properly installed
Compressor air is not connected
Air regulator is not set properly
Close front door
Replace can
Properly install the can
Connect air
Set regulator properly
Use without connecting handpiece
Absorption pads is saturated
Can is not properly installed
Air regulator is not set properly
Select only couplings that have a
handpiece connected
Clean TT and replace
filter
Properly install the can
Set regulator properly
Bent spray can nozzle pin
Wrong stand for spray can
Replace can
Use proper stand for can
Air pressure too high
Reduce air pressure, but not less
than 4 bar
No burr inserted in straight
attachment.
Refer to the user instructions for
the handpiece. For some
handpieces, an oil mist may be
released if a burr is not inserted
Oil mist is released.
Equipment surfaces get oily
Type of handpiece or coupling
Depending on the type of
handpiece and coupling, there
could be a release of oil mist. In
this case, lower the pressure to 4
bar
Plug in properly
*
Send the product every 2 years for a service check.
The safety checks in different countries can vary in compliance with country specific
regulations and requirements for medical devices. The national valid regulations are to be
observed.
*
Repair and maintenance work- part from the activities described in these operating
instructions-may be performed only by qualified technical staff.
*
Expected service life 5 years.
• In the event of modification by third parties, existing medical device licences become null and
void.
• Only use original parts and spare parts.
• The user( i.e. clinic, hospital etc.) is responsible for the management, maintenance and use of
medical devices.
■
Regular Inspection
*
Maintenance and inspection are generally considered to be the duty and obligation of the user, but
if, for some reason, the user is unable to carry out these duties, they may rely on a qualified
medical device serviceman. Contact your local dealer or the TTIBO CORP. for details.
•
Regularly Required Replacement Parts
Fuse
•
Other Replacement Parts
Absorption pads, Blister
(1) Power Supply Cord
Inspect visually for wear and broken wires.
(2) Main and Operation Switches
Turn on main switch and ensure main lamp lights up.
Check that the unit operates correctly.
(3) Absorption pads
Replace absorption pads(when the tray is filled by oil)..
En
glish