(2) Power Cord
Connect one end of the power cord to the main unit
and plug the other end into the electrical supply.
Power Cord
08
09
4. Setup
■
Note
Leave at least 5 cm (2 in) of free and open space around the TTcare 100+.
(1) Air Connection
1. Connect the air tube to the air connection
and the main air supply.
Turn on the valve for the main air supply.
■
Note
The main air supply must be clean and oil-free.
Make sure the tube is securely connected.
2. Set the air regulator dial between 4 and 6 bar.
Recommended Air Pressure is 4 bar.
■
Note
Less than 4 bar or higher than 6 bar will result in an
error code.
(3) Adapter
Fit the adapter into the nozzle and fasten the nut.
Nut
■
Note
Make sure the nut is securely tightened. Otherwise oil
and air will escape and cause incorrect maintenance.
2. Ensure the lock lever is on the left, place the spray
can on the stand shown in the photo.
Spray can
Stand
Unlock
3. Shift the lock lever to the right to secure the spray can.
Lock
■
Note
Ensure the nozzle is properly connected.
Make sure the lock lever is all the way to the right,
otherwise the oil will not come out.
Gently pull the spray can to ensure it is secured.
Oil will leak or disperse if the cans are not properly installed.
NOTE
(5) Spray Cans
1. Open the side door.
Press the PUSH sign indicated in the photo to open the
door.
PRESS
■
Note
Do not use Acidity Cleaner which is used for air
bearing handpieces only. It could damage both the
TTcare 100+ and the handpiece.
(4) Tray
Place two Absorption pads in the tray and slide it into
place.
Absorption
pad
Tray
■
Note
Slide the tray all the way in; otherwise, it might leak.