INTERFACE MENU
MENU ITEM
SOFTWARE
NAME
ITEM DESCRIPTION
ITEM RANGE
DEFAULT
VALUE
NETWORK
PROTOCOL
Premium
models only
NET
PROTOCOL
The NET PROTOCOL item selects the communications
protocol used to interface with the building management
system.
MODBUS
or
CIMETRICS
MODBUS
NETWORK
ADDRESS
Premium
models only
NET
ADDRESS
The NET ADDRESS item is used to select the main network
address of the individual room pressure device. Each unit on
the network must have its own unique address. The values
range from 1 to 247. If RS-485 communications are being used,
then a unique NET ADDRESS must be entered into the unit.
There is no priority between the RS-485 and keypad. The most
recent signal by either RS-485 or keypad will initiate a change.
RS-485 communications allows you access to all menu items
except calibration items. The RS-485 network can initiate a
change at any time. The RS-485 cannot override the TSI key
switch when ROOM MODE selection is key switch.
TSI key switch
A keylock switch mounted on the wall changes the pressure
between negative pressure, positive pressure, or no isolation (if
3 position switch installed).
1 to 247
1
OUTPUT
RANGE
OUTPUT
RANGE
The OUTPUT RANGE item selects the resolution range of the
analog pressure signal output. There are two choices:
LOW -0.01000 to +0.01000 "H
2
O.
HIGH -0.10000 to +0.10000 "H
2
O.
0 volt / 4 mA represents maximum negative pressure
differential.
5 volts / 12 mA represents a 0 room pressure differential.
10 volts / 20 mA represents maximum positive pressure
differential.
If the actual pressure exceeds the range, the output remains at
maximum or minimum depending on direction of flow.
LOW or HIGH
HIGH
Tech
ni
cal
Sectio
n
23
Summary of Contents for PRESSURA 8630-PM
Page 2: ......
Page 6: ...iv This page intentionally left blank...
Page 44: ...Specifications 38 This page intentionally left blank...
Page 45: ...Appendix B 39 Appendix B Wiring Diagram Figure 10 Wiring Diagram...
Page 46: ...Wiring Diagram 40 This page intentionally left blank...
Page 48: ...Access Codes 42 This page intentionally left blank...