CONFIGURE MENU
(continued)
MENU ITEM
SOFTWARE
NAME
ITEM DESCRIPTION
ITEM RANGE
DEFAULT
VALUE
SENSOR
SPAN
SENSOR
SPAN
The SENSOR SPAN item is used to match or calibrate the
PresSura velocity sensors to the average room pressure
velocity as measured by a portable air velocity meter.
A sensor zero should be established prior to adjusting the
sensor span (see
Calibration
section following menu item
listing).
NONE
Unit is factory
calibrated.
No initial
adjustment
should be
necessary.
FLOW ZERO
Premium
models only
FLOW
ZERO
The FLOW ZERO item is used to calibrate the PresSura flow
station pressure transducer.
A zero or no flow point needs to be established prior to using
the flow measurement (see
Calibration
section following menu
item listing).
NONE
Flow station
pressure
transducer
zero must be
done prior to
making any
flow
measuremen
ts.
SECOND
SENSOR
ZERO
Premium
models only
2SENSOR
ZERO
The 2SENSOR ZERO item is used to calibrate the second
PresSura pressure sensor.
A second sensor zero should be established prior to adjusting
the second sensor span. (see
Calibration
section following
menu item listing).
If menu item displays SEC SENS OFF, the second sensor must
be turned on (menu item 2 SENSOR ) to activate the
2SENSOR ZERO.
NONE
Unit is factory
calibrated.
No initial
adjustment
should be
necessary.
Tech
ni
cal
Sectio
n
21
Summary of Contents for PRESSURA 8630-PM
Page 2: ......
Page 6: ...iv This page intentionally left blank...
Page 44: ...Specifications 38 This page intentionally left blank...
Page 45: ...Appendix B 39 Appendix B Wiring Diagram Figure 10 Wiring Diagram...
Page 46: ...Wiring Diagram 40 This page intentionally left blank...
Page 48: ...Access Codes 42 This page intentionally left blank...