User Basics
5
Alarms
PresSura controller has visual (red light) and audible alarms to inform you of changing
conditions. The alarm levels (set points) are determined by hospital personnel, which
could be the infection control group, charge nurse, or a facilities group depending on the
hospital organization.
The alarms, audible and visual, will activate whenever the preset alarm level is reached.
Depending on the PresSura model installed, programmed alarms will activate when room
pressure is low or inadequate, when room pressure is high or too great, or when the
supply or exhaust air flow is insufficient. When the isolation room is operating safely, no
alarms will sound.
Example: The low alarm is preset to activate when the room pressure reaches -0.001
inches H
2
O. When the room pressure drops below -0.001 inches H
2
O (gets
closer to zero), the audible and visual alarms activate. The alarms turn off
(when set to unlatched) when the unit returns to the safe range which is
defined as negative pressure greater than -0.001 inches H
2
O.
Visual Alarm Operation
The red light on the front of the unit indicates an alarm condition. The red light is on for all
alarm conditions, low alarms, and high alarms.
Audible Alarms
The audible alarm is continuously on in all low and high alarm conditions. The audible
alarm can be temporarily silenced by pressing the
MUTE
key. The alarm will be silent for
a period of time (see MUTE TIMEOUT to program time period). When the timeout period
ends, the audible alarm turns back on if the alarm condition is still present.
You can program the audible alarm to be permanently turned off (see AUDIBLE ALM).
The red alarm light will still turn on in alarm conditions when audible alarm is turned off.
The audible and visual alarms can be programmed to either automatically turn off when
the unit returns to the safe range or to stay in alarm until the
RESET
key is pressed (See
ALARM RESET).
Summary of Contents for PRESSURA 8630-PM
Page 2: ......
Page 6: ...iv This page intentionally left blank...
Page 44: ...Specifications 38 This page intentionally left blank...
Page 45: ...Appendix B 39 Appendix B Wiring Diagram Figure 10 Wiring Diagram...
Page 46: ...Wiring Diagram 40 This page intentionally left blank...
Page 48: ...Access Codes 42 This page intentionally left blank...