42
Appendix A
Example 1:
Suppose your air flow temperature is 55ºF and you local barometric
pressure is 13.5 psia, and the standard flow indicated on the
A
CCU
B
ALANCE
is 200 (standard) CFM.
Actual
Flow
Actual CFM
=
⎡
⎣⎢
⎤
⎦⎥
=
200
212
14.7
13.5
460 + 55
530
Example 2:
Suppose your air flow temperature is 25ºC, and your local barometric
pressure is 730 mmHg, and the standard flow indicated on the
A
CCU
B
ALANCE
is 100 (standard) l/s.
Actual
105
Flow
Actual l s
=
+
⎡
⎣⎢
⎤
⎦⎥
=
100
760
730
273.15 25
294.25
✟