28
Model 8710 DP-C
ALC
TM
Micromanometer and Model 8375 A
CCU
B
ALANCE
®
Air Balancing Tool
4.
Press
X
until the display shows
SOURCE
U-F
for the desired measurement.
5.
Press
. The display shows
SOURCE
and whether
USEr
or
FACt
is selected for that measurement.
6.
Press
S
or
T
to select
USEr
or
FACt.
7.
Press
to save the selection.
Notes:
For the accessory temperature/humidity probe it is not possible to mix user and factory
calibrations for the different measurements. For example,
USEr
cannot be selected for
temperature with
FACt
selected for humidity.
The hood flow calibration factor is applied when the offset factor is changed. There is no
option to select
USEr
or
FACt
calibration for hood flow measurement. Set value to 1.000 to
restore factory calibration.
Connect and Download Data to an Optional Printer
To connect a printer, use the RS232 printer cable supplied with the optional printer. Note that the printer
interface cable is different from the computer interface cable.
The printer must be set to the same baud rate as the instrument. To change the baud rate of the printer,
refer to the printer’s operations manual. Always turn the instrument on
before
the printer. If the printer
prints question marks (??????), asterisks (******), or random characters, reset it by turning it off and then
on again. Figure 12 illustrates the printout information for a single sample.
---------------------------
MODEL: 8710 DP-Calc
SERIAL: 0404004
---------------------------
TEST ID: 2
Sample: 1
02/16/04
15:14:31 1.00 CFM
Shape=Round
Dia = 4.0 in. ø
Entered Temp= 74.4 °F
Baro. Press= 29.16 in.Hg
15:14:31 28.5 %rh
Figure 12: Sample Print Out
The Model 8710
allows printing of the entire memory, all Samples stored within a particular Test ID, or
an individual Sample.
To print the entire memory, press and hold the
key. This starts a countdown from 3 to 0 on the
display. When the display shows “0”, release the button. If you release the key at any time other than 0
during the countdown, nothing will print. The display will flash “Send dAtA” while printing.