Chapter 4: How To
19
Set the Time and Date
1.
Press
to display
CHANGE Test ID.
2.
Press
X
until
SETUP
appears.
3.
Press
.
4.
Press
X
until
TIME
is displayed.
5.
Press
. The display shows the current time and date. The hours value is flashing.
6.
Use
W
and
X
to move between values and
S
and
T
to adjust values.
7.
Press
when you have set the time and date as desired.
Note:
You can change the date format (see “
Set Format
” below). Allowable formats are mm.dd and
dd.mm.
Change the Currently Selected Test ID
To change the Test ID in which Samples will be stored, proceed as follows:
1.
Press
to display
CHANGE Test ID.
2.
Press
to display the current Test ID.
3.
Use
S
and
T
to select the Test ID you want to use.
Note:
If you press
S
again after the highest numbered Test ID is displayed, you will add a new Test
ID. (The Test ID is empty until you send Samples to it.) Also, if you press
SAVE
(or try to
automatically log data to a Test ID) and the Tool or display units or measurement
parameters are different from the information already stored in the currently selected Test
ID, the instrument will automatically change the currently selected Test ID to a new unused
Test ID. See the description of “
TEST IDs
” in Chapter 3.
Recall Sample Data
To recall Sample data so that you can review it or send it to a printer or computer, proceed as follows:
1.
Press
to display
CHANGE Test ID.
2.
Press
X
until
RECALL appears.
3.
Press
to display the current Test ID.
4.
Use
S
and
T
to select the Test ID you want to recall a Sample from.
5.
Press
. The
COUNT
for the selected Test ID is displayed. (Count is the number of Samples in the
Test ID.)
Press
S
or
T
to display an entry in the selected Test ID. In addition to displaying individual Sample
values, you can also view the MAX, MIN, AVG, and SUM of the data in the Test ID. (SUM
is only
available for Diffuser Flow [flow from pressure and K-Factor].)
Hold down
S
or
T
to scroll quickly
through the data.