26
Model 8710 DP-C
ALC
TM
Micromanometer and Model 8375 A
CCU
B
ALANCE
®
Air Balancing Tool
2.
Press
X
until
SETUP
appears.
3.
Press
.
4.
Press
X
until
BP.COMP
appears.
5.
Press
. The display shows whether back pressure compensation is on or off.
6.
Press
S
or
T
to change the option. Options are:
ON
and
OFF.
7.
Press
.
BP.COMP
is displayed.
8.
Press
ESC
to return to the main display.
Take Back Pressure Compensated Hood Readings
When back pressure compensation is turned on, the sequence of taking flow readings with the capture
hood is somewhat different.
1.
Open the flap in the capture hood base.
2.
Hold the capture hood to the diffuser being measured to capture its air flow.
3.
Press
READ
on the meter face or the red “Read Switch” on the left side of the hood base.
READ
OPEn
flashes on the display.
4.
When the open-flap reading is done,
CLOSE tHE FLAP
is displayed.
5.
Close the flap and press
READ
(or the Read Switch)
. READ CLOSEd
flashes on the display.
6.
When the closed-flap reading is done, the backpressure compensated flow is displayed.
7.
If you want to save the flow reading to log memory, press
SAVE
now.
STORE
is shown on the
display, then the saved reading is shown.
8.
Before pressing any more buttons, open the flap.
9.
If the display mode is set to
SINGLE,
pressing
READ
now will start a new backpressure
compensated reading.
If the display mode is set to
RUNAVG,
pressing
READ
now will measure and display a running
average of
non-backpressure-compensated
readings. Press
READ
once more to start a new
backpressure compensated reading.
Set Deadband On and Off
The deadband determines the velocity or volume measurement threshold for displaying 0. If the deadband
is
ON
, readings below 18 fpm (0.0914 m/s) or 18 cfm (30.58 m
3
/h, 8.49 l/s) will be displayed as 0. If the
deadband is
OFF
, readings down to 0 will be displayed. The default is set to
ON
.
1.
Press
to display
CHANGE
Test ID
.
2.
Press
X
until
SETUP
appears.
3.
Press
.
4.
Press
X
until
DEADbAnd
appears.
5.
Press
. The display shows whether
DEADbAnd
is on or off
.