12
EN
dehumidifier TTK 70 HEPA
Errors and faults
The device has been checked for proper functioning several
times during production. If malfunctions occur nonetheless,
check the device according to the following list.
The device does not start:
• Check the power connection.
• Check the power cable and mains plug for damage.
• Check the on-site fusing.
• Check the condensation tank for correct seating.
• Check the room temperature. Observe the device's
permissible operating range according to the technical
data.
• Wait for 10 minutes before restarting the device. If the
device is not starting, have the electrics checked by a
specialist company or by Trotec.
The device is running, but there is no formation of
condensate:
• Check the float in the condensation tank for dirt. If
necessary, clean the condensation tank. The float must be
able to move freely.
• Check the room temperature. Observe the device's
permissible operating range according to the technical
data.
• Ensure that the relative room humidity complies with the
technical data.
• Check the preselected desired room humidity level. The
relative humidity in the room must be above the selected
range. If necessary, reduce the preselected desired room
humidity.
• Check the air filter and HEPA filter for dirt. If necessary,
clean or change the air filter or HEPA filter.
• From the outside, check the condenser for dirt (see
chapter Maintenance). If the condenser is dirty, have it
cleaned by a specialist company or by Trotec.
The device is running, but there is no formation of
condensate:
• Check the float in the condensation tank for dirt. If
necessary, clean the condensation tank. The float must be
able to move freely.
• Check the room temperature. Observe the device's
permissible operating range according to the technical
data.
• Ensure that the relative room humidity complies with the
technical data.
• Check the preselected desired room humidity level. The
relative humidity in the room must be above the selected
range. If necessary, reduce the preselected desired room
humidity.
• Check the air filter and HEPA filter for dirt. If necessary,
clean or change the air filter or HEPA filter.
• From the outside, check the condenser for dirt (see
chapter Maintenance). If the condenser is dirty, have it
cleaned by a specialist company or by Trotec.
The device is loud or vibrates:
• Check whether the device is set up in a stable and upright
position.
Condensate is leaking:
• Check the device for leaks.
The compressor does not start:
• Check the room temperature. Observe the device's
permissible operating range according to the technical
data.
• Ensure that the relative room humidity complies with the
technical data.
• Check the preselected desired room humidity level. The
relative humidity in the room must be above the selected
range. If necessary, reduce the preselected desired room
humidity.
• Check whether the overheating protection of the
compressor has tripped. Disconnect the device from the
mains and let it cool down for approx. 10 minutes before
reconnecting it.
• The device might carry out an automatic defrost. During
automatic defrost, the device does not dehumidify.
The device gets very warm, is loud or is losing
performance:
• Check the air inlets and air filters for dirt. Remove external
dirt.
• From the outside, check the device for dirt (see chapter
Maintenance). If the inside of the device is dirty, have it
cleaned by a specialist company for cooling and air-
conditioning or by Trotec.
Note
Wait for at least 3 minutes after maintenance and
repair work. Only then switch the device back on.
The device still does not operate correctly after these
checks:
Please contact the customer service. If necessary, bring the
device to a specialist company for cooling and air-conditioning
or to Trotec for repair.