TX9165 User Manual
TX9165-UM-EN-07 87
Checkpoint
If there is a discrepancy of greater than 5%
of reading then recalibrate the appropriate
eModule of the Sentro 8 Sensorstation.
Or consider replacing the eModule.
Checkpoint
If there is a discrepancy of greater than 5%
of reading then consider changing the gas
sensing eModule.
Checkpoint
• Ensure that the appropriate Health and
Safety guidelines and applicable local
procedures are followed when handling
test gases.
• The ambient temperature should be
b20 and +30°C during the
checking procedure.
• Ensure the correct operation of the
Setpoint 1
and
Setpoint 2
alarms.
• Institute a formal checking and
maintenance plan.
• Operate a Maintenance and Calibration
Log and ensure it is updated every
time any Maintenance or Calibration
work is carried out.
Section 4.4.1
Section 7.4
Section 7.4