High
Air bubble in the infusion line
Press
【
Stop
】
button to stop
alarm, disconnect the line
from the patient, exhaust air
with air exhaust function, or
open the infusion pump door
to manually remove the air
bubbles
Door Open
High
During infusion, the infusion pump
door is opened
Close the infusion pump door
to stop this alarm.
System Error
High
Internal
failure
or
software
exception
Turn off and Restart, if the
alarm still exists, please
contact Triumph Medical
Note: When alarm rings, click the
『
Mute
』
icon on the screen to temporarily stop sound alarm for
2min.
Warranty
Warranty for both parts and labor is 12 Months from date of purchase. See extended warranty card
inside box for more information or visit Triumphmed.com.