16
●
Please confirm that the infusion line brand and specification displayed in the display screen is
accordant with the actual one in use.
●
If infusion accuracy is in question, the user is strongly suggested to contact Triumph Medical.
9) Install Drop sensor
Please install. After installing, click
『
Settings
』
→
『
Drop sensor
』
to activate the drop sensor
function.
Caution
●
The default state of drop sensor function is OFF, this function can be manually
activated by the user when the drop sensor is adopted.
Set Infusion Parameters
Purge Air
Under the parameters setting interface, Press
【
Bolus
】
button on product panel and hold on, or
touch the screen purge icon
to eliminate the air bubble in the line.
The purge total volume is not calculated in the Total Volume Infused.
Cautions:
●
Before purging air, double check to confirm the infusion line is
NOT CONNECTED
with the
patient.
●
Purge rate is set to the max flow rate, when purge volume
≥
5ml, purge will automatically stop.
Start Infusion
Connect IV tube with patient, confirm IV Line Brand, confirm infusion parameters, Press
【
Start
】
button
, click
『
yes
』
in the pop-up prompt interface, start infusion.