21
Drip mode
Under this mode, set the VTBI and drop rate, and the system will automatically calculate the
infusion flow rate and time. Drip Sensor not included.
Note:
●
The flow rate under drip mode is calculated according to the specification of the
current infusion line. Before adopting the drip mode, please confirm that the specification of the
current infusion line is accordant with the specification displayed in the interface title bar display.
If it is not accordant, please contact the equipment maintenance technician to modify. Otherwise,
it may cause serious deviation of flow rate.
Drug library mode
Under this mode, Select Drug, set the Weight (body weight), Conc. Unit (concentration unit),
Dose and VTBI (Volume to be infused), the speed will automatically be calculated according to
parameters. Drug Library Editable (see below in System Settings).
System Settings
Settings
Click
『
Settings
』
icon in the main interface to enter into parameters setting interface.
Drug Library
Select Drug Library in Settings. Options to select “None”, “Commonly used”, “Others”, or
“DurgLib maintenance”. Select a drug from “Commonly Used” or “Others” to view parameters
and have the drug name displayed on screen while infusing. (Note: if you are trying to utilize the
Drug Library mode then make that selection under the menu “Modes”.
To remove drug name from infusion running interface, select “None” in drug library.
To Edit Drugs go to Settings – Drug Library – DrugLib maintenance – Enter Password 8888 –
Edit/Add/Delete 2 different Groups of Drugs. Default Groups labeled “Commonly Used” and
“Others”. There are 32 total spots available for drugs, 16 in each group.
KVO Rate
Click
『
KVO rate
』
, input the numerical value, after confirming, click
『
OK. This is accessible in
Settings
Bolus Rate
Set the default Bolus rate under Settings.