background image

 

1.

 

Safety Instructions 

    Warnings 

 

Before  using,  please  check  the  equipment,  connect  the  power  cord  and  accessories  to  ensure  that  it 

works  normally  and  safely.  If  there’s  anything  abnormal,  immediately  stop  working  and  contact 

Triumph  Medical’s  service  department.  Additionally,  the  adhesion  or  intrusion  of  fluid/drug  may 

possibly cause the equipment to fault and malfunction. Therefore, please clean the equipment after use, 

and store it correctly.   

 

 

This equipment must be operated by trained professional medical care personnel.   

 

 

This equipment

 i

s not applicable to blood transfusion.   

 

 

It  is  not  allowed  to  put,  and  use  the  equipment,  in  the  environment  with  anesthetic  and  other 

inflammable or explosive articles to avoid fire or explosion.   

 

 

It  is  not  allowed  to  store  or  use  the  equipment  in  the  environment  with  active  chemical  gasses 

(including  gasses  for  disinfecting)  and  moist  environment  since  it  may  influence  the  inside 

components of the infusion pump and may possibly cause performance drop or damage of the inside 

components.   

 

 

The operator shall guarantee that the inputted infusion parameters of this equipment are the same as 

the medical advice before starting infusion.   

 

 

Please  correctly  install  the  infusion  line  according  to  the  infusion  indication  direction  of  this 

equipment,  ensure  that  the  infusion  tube  is  placed  smoothly  and  straightly  cross  the  creep  device. 

Otherwise, it may possibly suck blood from the patient or fail to reach the expected performance.   

 

 

Please  do  not  only  depend 

on  the  alarm  system  during  use.  Periodically  check  the  equipment  to 

avoid any accidents.

 

 

 

Tightly fix the equipment on the infusion stand/pole and ensure the stability of the infusion stand/pole. 

Be careful when moving the infusion stand and this equipment to avoid the equipment dropping and 

infusion stand/pole falling or knocking the surrounding objects.   

 

 

If the  infusion  tube is twisted, or the  filter  or needle are  obstructed,  or blood  is in the needle  which 

may obstruct the infusion, the pressure in the infusion tube will rise. When removing such occlusion, it 

may possibly cause “bolus injection” (temporary excess infusion) to the patient. The correct method is 

to tightly hold or clamp the infusion tube near the puncturing position, then open the door to drop the 

pressure in the infusion tube. Then loosen the infusion tube, solve the reason of occlusion, and restart 

infusion. If infusion is restarted when the occlusion reason exists, then it may cause occlusion alarm 

persistently,  and  the  pressure  in  the  infusion  tube  may  keep  rising,  and  may  break  or  cut  off  the 

infusion tube, or hurt the patient.   

 

 

This equipment injects fluid/drug through extruding the infusion tube, but it can’t detect the leakage if 

Summary of Contents for IP12

Page 1: ...User Manual Triumph IP12 Infusion Pump ...

Page 2: ...nts 1 Safety 3 4 2 Specifications 5 6 3 Appearance 7 11 4 Installation 12 13 5 Basic Operation 14 6 Infusion Operation 15 19 7 Settings 20 25 8 Alarms Troubleshooting 26 27 9 Maintenance Repair Warranty 28 31 ...

Page 3: ...ation direction of this equipment ensure that the infusion tube is placed smoothly and straightly cross the creep device Otherwise it may possibly suck blood from the patient or fail to reach the expected performance Please do not only depend on the alarm system during use Periodically check the equipment to avoid any accidents Tightly fix the equipment on the infusion stand pole and ensure the st...

Page 4: ...part may possibly have become reddish swelling or bleeding Additionally it is possible that the device doesn t alarm for a long period if the actual occlusion pressure is lower than the alarm threshold value therefore please periodically check the puncturing part If there s any abnormal phenomenon for the puncturing part please timely take suitable measures such as puncturing again Only the infusi...

Page 5: ...n Parameters Infusion line specification 10 60 drops System Accuracy 5 Drip Infusion Rate Accuracy 5 or 1drops min Infusion Rate 0 10 1500ml h Drip mode range 1 500drops min Bolus Rate 0 1 1500ml h Bolus preset 0 1 50ml KVO Rate 0 1 5 00ml h Micro mode setting range 100 1500ml h Minimum flow rate increment 0 01ml h VTBI 0 9999 99ml minimum step is 0 01ml Total Volume Infused 0 9999 99ml minimum st...

Page 6: ...e anesthetic gas mixed with air and inflammable anesthetic gas mixed with oxygen or nitrous oxide Operating 1 temperature 5 40 2 humidity 20 90 non condensable 3 atmospheric pressure 86 106kPa Transport Storage 1 temperature 20 60 2 humidity 10 95 non condensable 3 atmospheric pressure 50 106kPa Safety Standard Main Safety Standards IEC 60601 1 2005 A1 2012 Medical Electrical Equipment Part 1 Gene...

Page 7: ...7 3 Appearance Front View Tubing guide Pump door Pressure Plate Pressure sensor UPSTREAM if applicable Pump tablets Pressure sensor DOWNSTREAM Air in line sensor Anti free flow clamp ...

Page 8: ...election button AC indicator light When connecting with AC power supply AC indicator lights on Alarm indicator While pump alarms indicator light glitter different level different frequency and color more information please refer to Chapter 10 1 Running lights Start stop Bolus Purge Home Enter system home page ...

Page 9: ...con is WIFI indication icon Indicates WIFI connection state Battery charging indication icon Displays the current battery charging state Battery status indication icon The percentage numerical value at the left side of the icon displays the remaining battery Since the remaining battery may change it may possibly show the following states Typical Interface During pre infusion and infusion the typic...

Page 10: ... of the title bar input box editing box 1 Title bar displays the name of current editing parameter 2 Input box real time display of the input content 3 Editing box It composes of the main button area and function button area The main button area is composed of the numerical values and letters and icons The function button area is composed of the clear button cancel button and Shift Icon Paraphrase...

Page 11: ... Port External 12V DC power supply Handle A C Adapter Port External 100 240V 50 60Hz AC power supply Loudspeaker Pole Clamp Using for fixing the equipment on the infusion stand IrDA Using for communicating with infusion docking station Optional Latch for stackable function ...

Page 12: ...em to handle the packing materials Installation Warning This equipment shall be installed by qualified personnel All devices that connect with this equipment must pass the designated IEC standards for example IEC60950 information technology equipment safety and IEC60601 1 medical electric device safety certification and all devices must be connected according to the valid version of IEC60601 1 1 s...

Page 13: ...ole rotate the pole clamp screw knob and unscrew to leave space Lock the Pole Clamp on the infusion stand adjust the position of the infusion pump tighten the pole clamp to fix the infusion pump on the infusion stand shown in drawing to the right Hold the infusion pump when tightening the fixing clamp loose it after tightening to avoid falling 4 The pole clamp supports the vertical pole at default...

Page 14: ...IV set brand name Select infusion tube brand Remove air bubble from the line Select infusion mode Select infusion modes according to requirement Set infusion parameters set infusion parameters according to requirement Connect the infusion line with the patient Start infusion press start infusion Infusion finish Remove the IV Set Power off or Standby ...

Page 15: ...ag 2 Extrude the drip chamber when the fluid has reached 1 2 position of the drip chamber open the roller clamp 3 Fill fluid drug to the injection needle to remove air then close the roller clamp 4 Pull the lock switch door handle in the middle of the pump door from the lower side then open the door 5 Push up anti free flow clamp to open it 6 Install the infusion tube in the infusion tube slot acc...

Page 16: ...r when the drop sensor is adopted Set Infusion Parameters Purge Air Under the parameters setting interface Press Bolus button on product panel and hold on or touch the screen purge icon to eliminate the air bubble in the line The purge total volume is not calculated in the Total Volume Infused Cautions Before purging air double check to confirm the infusion line is NOT CONNECTED with the patient P...

Page 17: ...p will go back to the previous setting infusion rate 2 Automatic bolus Under the running interface click Bolus on touch screen set two parameters among bolus infusion volume rate and time click Start It will beep at every 1ml infused After bolus infusion is finished the equipment goes back to the previous infusion rate Infusion Completion When remaining infusion time is near preset volume to be in...

Page 18: ...a standby function available while not infusing Replace Infusion Line Infusion Container Please replace the infusion tube assembly according to the following steps Close the flow rate adjuster of the infusion tube assembly open the infusion pump door and then remove the infusion tube assembly Prefill and install the new infusion tube assembly Operate to restart infusion according to the above infu...

Page 19: ...For both the rate set in infusion parameter and the rate calculated by the system the range is the system default flow rate of the current working infusion line specification 4 If no VTBI Volume to be infused is Set then the fluid drug in the infusion container will be completely used Infusion Parameters Setting Range Infusion Parameter Parameter Range VTBI 0 9999 99ml Rate 0 1 1500ml h Time 1min ...

Page 20: ... the third parameter If volume to be infused is 0 pump will work until pumps alarms Time Mode Under this mode it allows to set VTBI Volume to be infused and time The system will automatically calculate the speed speed Volume ml time min Body Weight Mode Under this mode set the Weight body weight Active agent drug mass Conc unit concentration unit Volume fluid volume Dose rate Dose unit VTBI The sy...

Page 21: ...rameters Drug Library Editable see below in System Settings System Settings Settings Click Settings icon in the main interface to enter into parameters setting interface Drug Library Select Drug Library in Settings Options to select None Commonly used Others or DurgLib maintenance Select a drug from Commonly Used or Others to view parameters and have the drug name displayed on screen while infusin...

Page 22: ...he system will report occlusion alarm even when the line is not obstructed Under this condition please carefully observe the pressure indication icon in the display screen and infusion line and raise the occlusion pressure if needed When the occlusion pressure is set at a high level it may possibly cause the patient discomfort After raising the occlusion pressure please carefully observe the condi...

Page 23: ...rding to the actual requirement Air Bubble detector level Alarm Threshold Value Level 1 50ul Level 2 100ul Level 3 200ul Level 4 300ul Level 5 450ul Level 6 600ul Level 7 800ul Reminder Alarm Click Reminder alarm to enter into the time for reminder alarm setting interface click the preset time option to set the reminder alarm time Finish Pre Alarm Time for pre alarm refers to the time of activatin...

Page 24: ...prompt box click Yes to confirm reset otherwise please click No General In the main interface click General to enter into the General equipment setting interface Date Time Click Date Time to enter into the date and time setting interface It allows to set the date time and format in this interface When setting date and time directly input the numerical value in the input method interface For exampl...

Page 25: ...s not touched Unlock Click Cancel in the lock screen interface Note The equipment will automatically unlock if there s a high Level alarm Night Mode Click Night mode to enter into night mode Switch setting interface to set the start and end time of the night mode and the night brightness At night the system automatically adjusts the brightness to the user defined value Battery Capacity Display Tur...

Page 26: ...ow click OK the sound signal will stop and exit the alarm interface Click Mute to mute if alarm is not eliminated the alarm sound will reoccur in 2 minutes Warning Some alarm threshold values of this equipment can be set by the user For example occlusion pressure air bubble alarm reminder alarm VTBI infused pre alarm alarm sound volume and so on The user shall confirm the parameters when set the a...

Page 27: ...o the prompt If it can t be eliminated please stop the equipment contact Triumph Medical to repair and test the equipment Do not put it into operation before the equipment has passed the inspection otherwise it may possibly cause unpredictable harm if it continues to operate under fault contition If the equipment is on fire burns for unknown reason or has other abnormal conditions the user shall i...

Page 28: ... drug may corrode the pump and cause working faults After infusion please timely clean the equipment wipe it with a moist and clean soft fabric and then naturally dry it 2 When cleaning the equipment interface please wipe it with dry and soft fabric Confirm the interface is dry before using 3 Please do not soak the equipment in water Although this equipment has certain waterproof function when flu...

Page 29: ...infusion Completion Press Stop button to stop alarm Pressure high High 1 Line occlusion during infusion Manually remove the reason of occlusion Press Start button to continue infusion 2 Fluid drug in the actual infusion line has high viscosity but the system occlusion level is set too low Rise the alarm Level Press Start button to continue infusion 3 The pressure sensor is damaged Please contact T...

Page 30: ...r supply High Battery is removed and the AC power wire is dropped Connect power supply Contact Triumph Medical Reminder alarm Low After installing infusion tube under non working or alarm state it is not operated within the set time of the system Click any button to stop Standby time expired Middle During standby after reaching the standby time Press Stop button to stop alarm KVO finished High KVO...

Page 31: ...During infusion the infusion pump door is opened Close the infusion pump door to stop this alarm System Error High Internal failure or software exception Turn off and Restart if the alarm still exists please contact Triumph Medical Note When alarm rings click the Mute icon on the screen to temporarily stop sound alarm for 2min Warranty Warranty for both parts and labor is 12 Months from date of pu...

Page 32: ...32 ...

Reviews: