18
Stop Infusion
During infusion or after infusion, click
, infusion stop. It will return to the parameter setting
interface display Total Volume Infused and adjustable parameters.
Remove the Infusion Line
Disconnect the IV set’s extension line from the patient, then remove the IV set.
Replace IV set
Power OFF or Standby
Method 1: hold the
【
Power
】
Button till the screen is OFF, the equipment is OFF.
Method 2: press the
【
Power
】
Button to enter into OFF interface.
(1) Turn off the equipment: click
『
Power off
』
icon, the equipment is turned OFF.
(2) Standby: click
『
Standby
』
icon to enter into standby time setting interface, set the standby time.
Under standby state, the screen brightness will be lowest, after standby, the screen brightness will
be recovered.
(3) Cancel: click
『
Cancel
』
, return to the interface before OFF setting.
Note:
●
The equipment has a standby function available while not infusing.
Replace Infusion Line/Infusion Container
•
Please replace the infusion tube assembly according to the following steps:
- Close the flow rate adjuster of the infusion tube assembly, open the infusion pump door, and then
remove the infusion tube assembly.
- Prefill and install the new infusion tube assembly.
- Operate to restart infusion according to the above infusion steps if needed.
•
Please replace the fluid/drug container according to the following steps:
- Close the flow rate adjuster of the infusion tube assembly.
- Remove the fluid/drug container from the infusion tube assembly.
- Connect the infusion tube with the new fluid/drug container.
- Restart infusion according to the above steps of replacing infusion tube assembly.
Warning:
●
The infusion tube will distort if it is used for a long period of time and could
result in inaccuracy or flow rate error, it is suggested to replace the pumping position or infusion
tube assembly after working for 8 hours or before depending on flow rate.