Tritex Multigauge 5700 Datalogger Ultrasonic Thickness Gauge www.tritexndt.com
Page 19
6.9.
Setting the Resolution
Either 0.1 mm (0.002”) or 0.05 mm (0.005”) resolution can be selected.
i
Navigate to ‘Resolution’ on the ‘Home’ screen using the ‘Arrow’
keys (2,5,7 & 9).
Use the Definition of Menu Items on pg 13 as a
guide.
ii
Press ‘OK’ (4).
iii
Use the ‘Up’ (5) and ‘Down’ (9) buttons to select the required
resolution between ‘0.05’ and ‘0.1’ (metric) or ‘0.005’ and 0.002’
(imperial).
iv
Press ‘Option Key 2’ (6), the gauge will save the new resolution
and return to the ‘Home’ screen.
v
Pressing the ‘Option Key 1’ (1) will quit the resolution setup and
return to the ‘Home’ screen.
6.10. Coating Plus
+
Coating Plus
+
mode is used to measure metal thickness when thick coatings
up to 20 mm exist. Normal measurement mode will ignore coatings up to
approximately 6 mm. In both modes, the actual thickness of coating that can
be ignored depends on the type of coating.
Note:
Coating Plus
+
mode should not be left on for all measurements. The
gauge will achieve a better performance on standard coatings with this
function turned off.
SAVE
QUIT
Select Resolution:
ü
0.1 mm
0.05 mm
change Reading Resolution: