![TRISTAN TECHNOLOGIES SMM-770 Operating Manual Download Page 74](http://html1.mh-extra.com/html/tristan-technologies/smm-770/smm-770_operating-manual_1168331074.webp)
Model SMM 770 Scanning Magnetic Microscope System
TRISTAN TECHNOLOGIES --Page
74
Figure 7-3
Subtracted background
7.5 Stage calibration/test
X axis:
Minimum travel distance = -77 mm
13 May 05
CGA
Maximum travel distance = +77 mm
13 May 05
CGA
Backlash = 8 microns
27 May 05
CGA
Movement resolution < 10 micron
27 May 05
CGA
Nominal 500 micron movement = 19.5 mil 27 May 05
CGA
Nominal 50 mm movement = 50.02 mm
27 May 05
CGA
Y axis:
Minimum travel distance = -77 mm
13 May 05
CGA
Maximum travel distance = +77 mm
13 May 05
CGA
Backlash = 8 microns
27 May 05
CGA
Movement resolution < 10 micron
27 May 05
CGA
Nominal 500 micron movement = 19.5 mil 27 May 05
CGA
Nominal 50 mm movement = 49.99 mm
27 May 05
CGA
Z axis:
Minimum travel distance = 0 mm
27 May 05
CGA
Maximum travel distance = 25.4 mm
27 May 05
CGA
Backlash < 2 microns
27 May 05
CGA
X tilt:
Full 360 deg turn of knob, across 150 mm
left +17 mil, right -5 mil, 0.1165 deg
27 May 05
CGA
Y tilt:
Full 360 deg turn of knob, across 150 mm
back +17 mil, front -5 mil, 0.1165 deg
27 May 05
CGA