21
Triplett Model 9310-A
Turn the equipment or voltage source on and take the required reading. DO
NOT HANDLE OR TOUCH THE METER, TEST LEADS, OR ALLIGATOR
CLIPS WHILE THE EQUIPMENT IS TURNED ON.
8.2.2 If you ARE using a Lineman’s Glove
Touch the test prod tips to the circuit being measured and take the required
reading. The supplied alligator clips may also be used to attach the leads to
the circuit.
8.2.3
If the voltage is such that a different range on the Meter would be
appropriate, then:
If you ARE NOT using a Lineman’s Glove
Turn the power to the equipment off and allow the reading to return to zero.
Disconnect the leads from the equipment.
8.2.4 If you ARE using a Lineman’s Glove
Disconnnect the leads from the equipment.
Only after the leads are disconnected from the equipment should the range
be changed. After the new range is selected, reconnect the leads as
described above.
8.3 High Current Measurements
CAUTION:
Always observe the following rules and procedures when making
measurements in high current circuits. Never use the Meter on circuits
with voltages higher than 600VRMS.
Read and follow the rules for high voltage in Section 8.2.
Disconnect all test leads from the Meter.
Always use the Lever to open and close the Clamp Jaws. Never place
any part of your body beyond the Hand Guard.
Be careful when trying to clamp the meter around the conductor. To
properly place the Meter may require the hand holding the meter to
come close to other circuitry which could prove a serious shock hazard.
WARNING!!!
WHENEVER TOUCHING THE METER IN AN ENERGIZED CIRCUIT, THE USE
OF A LINEMAN’S GLOVE CONFORMING TO ASTM D120 REQUIRED.