18
Notes:
a)
The 2M, 20M, and 200M ranges require several seconds to stabilize.
b)
The 200M range reads 10 LSD high. To obtain an accurate reading, subtract
10 LSD from the indicated value on the LCD.
c)
To obtain the most accurate reading on the 200
Ω
range, short the test leads
together and note the “residual resistance” reading. It is typically less than
0.5
Ω
. Subtract this value from subsequent readings on the 200
Ω
range.
7.8 Continuity Beeper:
WARNING!!!
Do not apply voltage or current to the meter when it is set
to the Diode Test / Continuity Beeper “
” range.
7.8.1
Connect the black test lead to the
COM
jack and the red test lead to the
VOLT / OHM / BEEPER / DIODE / FREQ.
“
V
Ω
kHz
” jack.
7.8.2
Set the RANGE switch to the
Diode Test / Continuity Beeper
“
” range.
7.8.3
Connect the test probes to the device or circuit to be tested.
7.8.4
If the resistance of the device or circuit is less than 20 to 100 Ohms, the
continuity beeper will sound.
Note:
The reading displayed on the LCD is not an accurate indication of the
resistance of the device or circuit being measured.
7.9 Diode Test:
WARNING!!!
Do not apply voltage or current to the meter when it is set
to the Diode Test / Continuity Beeper “
” range.
7.9.1
Connect the black test lead to the
COM
jack and the red test lead to the
VOLT / OHM / BEEPER / DIODE / FREQ.
“
V
Ω
kHz
” jack.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.
Downloaded from
Arrow.com.