OSCAR G2
Manual
D01-44en201810
Software V.1.1.6
(C)TriOS Mess- und Datentechnik GmbH
Bürgermeister-Brötje-Str. 25
D- 26180 Rastede/Germany
25 / 62
Step 2) Plug in the M12 connector on the cable end of the
OSCAR
sensor to the M12 socket
(2) of the G2 interface box and close the screw cap.
Step 3) Connect the 12 or 24 VDC power supply of the
G2 Interface Box
to connect the
sensor with voltage.
Step 4) Wait at least 3 seconds before you finally connect the Ethernet LAN cable to your
Ethernet enabled device and the
G2 Interface Box
.
The web interface can be used with any web browser via the URL http://oscar/ or
http://oscar_B0XX/ (B0XX is the serial number) or http://192.168.77.1/
NOTE:
If the web interface is not accessible, make sure that the LAN cable is connected
after the sensor has been powered on and try all three URL options.
3.2 Mechanical installation
3.2.1 Submerged use
For submerged use, the
OSCAR
can be completely or partially immersed in the
measurement medium. The cavity has to be submerged completely. Use the mounting bolt
together with a shackle and a stainless steel chain or wire to hang the sensor in the fluid. Do
not carry or pull the sensor with the cable. It is also possible to use special clamps similar to
the ones shown in Fig. 7 to install the sensor.