Fig. 7. Clamps for installation of the
OSCAR
Take care to use matching clamps with an inner diameter of 68 mm. Use two clamps at the
sensor housing, one next to the cavity and one at the bottom. Take care, that the clamps are
mounted near the screws and the caps of the sensor to avoid damages by squeezing the
sensor tube. The clamps can be purchased from
TriOS Mess- und Datentechnik GmbH.
The operating pressure inside the cavity should not exceed 1 bar
difference compared to its surrounding medium. Make sure that the
sensor is installed in the correct position to ensure free water flow.
3.2.2 Operation on-site
The
OSCAR
can also be filled in an on-site operation. There is a specific holder supplied
with the
OSCAR
applicable as a stand in the laboratory (see Fig. 8).
In such an installation the inlet should be positioned at the bottom, the outlet on the top of the
cavity to reduce air bubbles.