facilitate the cleaning of the optical windows
TriOS Mess- und Datentechnik GmbH
offers a
cleaning set with acetone and special optical cleaning paper.
5.2 Cavity cleaning
The PSICAM cavity wall can be cleaned by using pure HPLC-grade ethanol and a lint-free
tissue. Touching the wall with unprotected finger should be avoided. Optical influences by
organic material attached to the wall can further be removed by bleaching for about 15
minutes using a ca. 0.1 % sodium hypochlorite solution (NaOCl). The central light source can
either be carefully cleaned with ethanol.
How to open the cavity?
Position the device horizontally and make sure it cannot slip away (Fig. 33). Even though the
top cap of the cavity fits in any orientation on the main part, we recommend making a small
mark with a pencil. The cavity can be opened with the four screws on the top cap.
Fig. 33.
OSCAR
positioned for cavity opening