Select the path to the appropria
te calibration file in the “File” field by clicking on the “Browse”
button. Then, click on the "Upload" button to start the transfer. If the operation is successful,
this will be indicated by a green box that reads "Success". Otherwise, you should see a box
with an error message at the top of the screen.
The following error messages and warnings are possible:
File not OK.
The calibration file could not be read correctly. Check the path and
select the correct file. If the error persists, contact
TriOS Mess- und Datentechnik
GmbH
customer service.
Device type or serial number does not match.
The calibration file is not suitable for
the currently connected sensor.
4.1.6 Calibration
This chapter is only an overview of G2 Interface. A detailed step by step description is
located in chapter 6.
On the page "Calibration"
OSCAR
can be calibrated. That means a new waterbase (zero
line) can be recorded and Nigrosin for the reflectivity calculation can be measured (Fig. 22).