1.4
MANIFOLD INSTALLATION
CAUTION: Handle the manifold carefully. The dispense and aspirate needles on the
manifold are fragile.
The tubing must be connected correctly for proper functioning of the manifold.
The
connectors are color coded to assist in correctly connecting the manifold to the washer tubing.
Connect the white dispense fittings together and the clear aspirate fittings together. Tubing
should look like the illustration when it is correctly installed.
If the manifold tubing is to be replaced, see section 4.8 to properly reposition it.
Place the soaking microstrip well into the slot as shown below.
Multiwash III
SINGLE MICROSTRIP WELL
USED WITH 8 AND 12-PORT STYLE MANIFOLDS
CLEAR FITTING
WHITE FITTINGS
CLEAR FITTING
SOAKER MICROSTRIP WELL
NEAREST TO PLATE
MANIFOLD
PLATE
4