4.7
REPLACING STANDARD MANIFOLD DISPENSE NEEDLES
Note: A needle service tool—included with the optional Maintenance Kit and in the
Replacement Needle Kit—is required to replace manifold dispense needles.
Insert the
needle service tool under the dispense needle collar. Pry the dispense needle upward with the
tool. This will loosen the needle from its press fit with the manifold. Place the new needle into
the open end of the needle service tool and firmly insert it into the manifold.
Care should be exercised to prevent breakage of the precision glass insert.
Multiwash III
26