SYMPTOM
PROBABLE CAUSE
REMEDY
11. Grinding noise from the back of
the instrument and no fluid
movement.
Syringe plunger is stuck due to
dried wash buffer freezing the
plunger to the syringe.
Turn the power Off and On and
verify if syringe is working.
Turn the power Off. Take a large
standard screw driver and turn the
Stepper Motor Nut, counter
clockwise 5 turns.
Turn the Power On.
The Syringe assembly needs to be
replaced with replacement, Catalog
# 938. Follow the directions
supplied with the replacement kit.
Syringe plunger is out of follower.
See symptom #13.
12. Syringe plunger pops out of
traveler nut.
Kink or blockage in tubing from
Wash and Rinse reservoir, to the
syringe valve.
Clear inlet tubing of all obstructions.
Reset syringe plunger.
Wash and rinse reservoir inlet filter
clogged.
Backflush or replace filters.
The syringe motor assembly needs
to be replaced. Contact your
distributor for assistance.
See Appendix 2.
13. “E02” displayed.
Syringe plunger has become
disconnected, or jammed. See
symptoms #11 and 12 for possible
causes.
Turn off the instrument to reset it
and see symptoms #11 and 12 for
remedies.
14. Any ERROR code other than
“E02”.
Damage to the instrument by
electrical outage, brown out, low
voltage, or high voltage spikes.
Turn off instrument to reset it and
contact your distributor for
assistance.
See Appendix 2.
5.2
TROUBLESHOOTING TABLE (CONTINUED)
Multiwash III
33
Summary of Contents for 8441 series
Page 1: ...8618 01 E MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL ...
Page 5: ...Multiwash III THIS PAGE INTENTIONALLY LEFT BLANK vi ...
Page 14: ... FINAL ASPIRATION WELL POSITION TOP OF WELL ASPIRATION POSITION Multiwash III 9 ...
Page 40: ...5 4 USER TROUBLE SHOOTING NOTES Multiwash III 35 ...